Caution
Make sure the cuff is wrapped in the correct arm position and is at the same height as the heart.
A difference of 10 cm (4 inches) in height may cause a variation in the blood pressure value of up to
7 - 8 mmHg.
Note:
• If measurement is difficult due to arrhythmia, use a different blood pressure measurement method.
• If the patient has acute inflammation, a pyogenic ailment, or an external wound at the location where
the cuff is to be wrapped, follow the instructions of the doctor.
• Non-Invasive Blood Pressure (NIBP) measurem
ent is performed by compressing the upper arm.
Some people may experience intense pain, or transient spotting caused by subcutaneous hemorrhaging
may appear. The spotting will disappear over time, however, if it is possible that this will disturb the patient,
try the following technique:
- Wrap a thin towel or cloth (one layer) under the cuff.
If the towel or cloth is too thick, there will be insufficient cuff compression and the blood pressure value
will measure high.
• If the patient moves or the cuff is touched, this
may be falsely detected as a pulse and over-inflation
will occur.
• Do not inflate the cuff when it is not wrapped on the upper arm. This may damage the cuff.
Taking the Measurement in “Manual Mode”
1. Press the [START/STOP] button.
Blood pressure measurem
ent is performed once.
2. The measurement results are displayed.
If a measurement value is outside the corresponding range below, the value will flash.
SYS: 59 mmHg or less, or 251 mmHg or higher.
DIA: 39 mmHg or less, or 201 mmHg or higher.
PULSE: 39 bpm or less, or 201 bpm or higher.
Caution
The accuracy of a flashing measurement value that is outside the measurement range is not
guaranteed. Always check the patient’s condition before deciding what steps to take.
Note:
If inflation is insufficient, inflation may restar
t automatically while measurement is in progress.
Q
Normal measurement
Q
Measurement error /
failure
HBP-1300-E main.book Page 21 Thursday, February 11, 2016 11:39 AM
Summary of Contents for HBP-1300
Page 232: ...230 HBP 1300 E main book Page 230 Thursday February 11 2016 11 39 AM...
Page 233: ...231 RU OMRON HBP 1300 E main book Page 231 Thursday February 11 2016 11 39 AM...
Page 234: ...232 IEC 60601 1 2 OMRON HBP 1300 E main book Page 232 Thursday February 11 2016 11 39 AM...
Page 235: ...233 RU SpO2 HBP 1300 E main book Page 233 Thursday February 11 2016 11 39 AM...
Page 236: ...234 HBP 1300 E main book Page 234 Thursday February 11 2016 11 39 AM...
Page 237: ...235 RU 6 20 30 C 4 86 F 65 20 HBP 1300 E main book Page 235 Thursday February 11 2016 11 39 AM...
Page 239: ...237 RU 241 252 OMRON 5 HBP 1300 E main book Page 237 Thursday February 11 2016 11 39 AM...
Page 244: ...242 1 2 3 4 5 4 300 4 HBP 1300 E main book Page 242 Thursday February 11 2016 11 39 AM...
Page 253: ...251 RU START STOP 3 HBP 1300 E main book Page 251 Thursday February 11 2016 11 39 AM...
Page 257: ...255 RU 30 HBP 1300 E main book Page 255 Thursday February 11 2016 11 39 AM...
Page 264: ...262 HBP 1300 E main book Page 262 Thursday February 11 2016 11 39 AM...
Page 308: ...HBP 1300 HBP 1300 HBP 1300 HBP 1300 d 1 2 P d 1 2 P d 2 3 P d P...
Page 311: ...AR EMC EMC EN60601 1 2 2007 EN60601 1 2 2007 OMRON HEALTHCARE EN60601 1 2 2007...
Page 312: ...NIBP EN1060 1 1995 A2 2009 EN1060 3 1997 A2 2009 ISO81060 1 2007 K5 93 42 EEC EN 1060...
Page 314: ...E3 OMRON E40 E41 E42 OMRON ABS...
Page 315: ...AR E40 E2 E9 OMRON NIBP E1 OMRON GS CUFF E2...
Page 316: ......
Page 317: ...AR OMRON OMRON OMRON START STOP NIBP...
Page 318: ...START STOP NIBP OMRON GS CUFF...
Page 319: ...AR HBP 1300 80 v v 50 v v NIBP 70 v v 80 v v...
Page 320: ...START STOP...
Page 321: ...AR ON MODE MODE...
Page 322: ...START STOP...
Page 323: ...AR INDEX ARTERY RANGE INDEX ARTERY RANGE INDEX ARTERY...
Page 324: ...GS CUFF XL GS CUFF L GS CUFF M GS CUFF S GS CUFF SS NIBP OMRON GS CUFF CUFF HOSE CUFF HOSE...
Page 326: ...START STOP MODE E40 MODE START STOP...
Page 327: ...AR MODE 280 AUTO 220 250 280 220 AUTO 280 250 220 AUTO MODE ON MODE START STOP OFF AUTO ON OFF...
Page 328: ...LCD LCD E40 E40 OMRON...
Page 329: ...AR...
Page 330: ...LCD...
Page 331: ...AR START STOP MODE NIBP BF...
Page 333: ...AR OMRON M GS CUFF L HXA BAT 2000 9065797 0...
Page 334: ...OMRON NIBP...
Page 335: ...AR NIBP luer IBP SpO2 OMRON GS CUFF...
Page 336: ......
Page 337: ...AR...
Page 338: ...SpO2...
Page 339: ...AR IEC60601 1 2 OMRON...
Page 340: ...OMRON MRI LCD...
Page 341: ...AR...
Page 342: ...OMRON OMRON OMRON OMRON OMRON OMRON OMRON OMRON...
Page 343: ...NIBP HBP 1300 AR...