5000 Series Indicators
.11. External Input
Set the functon to be controlled by an optonal external nput devce such as a foot swtch.
OFF = dsabled.
tArE = Tare functon.
ZErO = Zero functon.
PrINt = Prnt functon.
FUNCt = acton specfc to the current applcaton mode.
S-S = the frst external nput changes the state of the relay. The second external nput
(Start-Stop) returns the relay to the orgnal state.
t-S-S = the frst external nput ntates a Tare functon, the second external nput
(Tare-Start-Stop) changes the state of the relay. The thrd external nput returns the relay to ts
orgnal state.
.11. Input Beep
Set the beeper response to an external nput.
OFF = Dsabled.
ON = Enabled.
.11. Relay Output
Set the relay output parameters.
NOTE
: If the Relay opton s not nstalled the OUTPUT menu and assocated menu tems are not avalable.
Type
Set the ntal state of the relay.
OPEN = the relay output s normally open.
CLOSEd = the relay output s normally closed.
CAUTION
: The normally closed relay condton s only actve whle the Indcator s powered on.
When powered off or when power s removed, the relay condton returns to a normally open
condton. Restorng power to the Indcator wll restore the closed condton of the relays.
Output Sequence
Set how the relay outputs react as the weght readng changes from under / accept / over.
NOrM = the prevously enabled relay wll be dsabled as the next relay s enabled.
HOLd = the prevously enabled relay wll hold the same state as the next relay s enabled.