5000 Series Indicators
..1 Zero Calibration
Zero calbraton uses one calbraton pont. The zero calbraton pont s establshed wth no weght on the
scale. Use ths calbraton method to adjust for a dfferent pre-load wthout affectng the span or lnearty
calbraton. When ZErO s dsplayed, press the
Yes
button to ntate Zero Calbraton.
The dsplay flashes 0 and the calbraton unt. Press the
Yes
button to establsh the zero pont.
The dsplay shows --C-- whle the zero pont s establshed.
When zero calbraton s completed, the dsplay shows dONE.
Then the scale exts to the actve weghng mode and dsplays the actual weght value.
.. Span Calibration
Span Calbraton uses two ponts to adjust the scale. The span calbraton pont s establshed wth a
calbraton mass placed on the scale. The zero calbraton pont s establshed wth no weght on the scale.
When SPAN s dsplayed, press the
Yes
button to ntate Span Calbraton.
The dsplay flashes the span calbraton pont. Place the specfed weght on the scale and press the
Yes
button.
To choose a dfferent span pont or calbraton unt, edt the settng as explaned n Secton 3.2 Menu
Navgaton. When the desred settng s dsplayed, place the specfed weght on the scale and press the
Yes
button.
The dsplay shows --C-- whle the span pont s establshed.
The dsplay flashes 0.
Wth no weght on the scale, press the
Yes
button to establsh the zero pont.
The dsplay shows --C-- whle the zero pont s establshed.
When span calbraton s completed, the dsplay shows dONE.
Then the scale exts to the actve weghng mode and dsplays the actual weght value.