..5 Geographical Adjustment Factor
Refer to Table 3-2 and set the GEO factor that corresponds to your locaton.
00 to 31
NOTE:
Only an authorzed manufacturer’s representatve or certfed verfcaton personnel may make these
changes. Changng the geographcal settng alters the calbraton values.
Advance to the next menu.
.. End Calibration
.. Calibration Test
Calbraton test s used to compare a known calbraton weght aganst the stored span calbraton data.
NOTE
: Calbraton Test s always avalable (even when LFT s set to ON).
When tESt s dsplayed, press the
Yes
button to ntate Calbraton Test.
The dsplay flashes 0. Wth no weght on the scale, press the
Yes
button to record the current zero pont.
The dsplay shows --t-- whle the zero pont s recorded.
The dsplay flashes the span calbraton weght usng the value from the last calbraton. The example shows
test weght of 30 kg.
Place the specfed test weght on the scale and press the
Yes
button.
The dsplay shows --t-- whle the data s processed.
The dsplay flashes the actual dfference between the calbraton data and the test weght.
The example shows a 0.010 kg dfference. The result of the Calbraton Test s prnted.
After 5 seconds, Calbraton Test ends, the scale returns to the actve weghng mode and dsplays the current
weght.
•
•
•