EN-0
5000 Series Indicators
Reset
No
, Yes
Range
Single
, Dual
Full Scale Capacty
1
…999950
Graduaton
0.00001
…1000
Power On unt
Auto,
kg, g, lb, oz, lb:oz
Zero Range 2%,
100%
Auto-Tare
Off
, On, Accept
Retan Weght Data
Off
, On
Legal for Trade
Off
, On
Beeper Volume Off,
Lo
, H
Beeper Sgnal
Off
, Accept, Under, Over, Under- Over
Button Beep
Off
, On
End Setup Ext SETUP menu
. Setup Menu
When the Indcator s used for the frst tme, enter ths
menu to set the Range, Capacty and Graduaton. Default
settngs are
bold
.
..1 Reset
Reset the Setup menu to the factory defaults. (except Range, Capacty and Graduaton)
NO = not reset.
YES = reset.
NOTE
: If the Legal for Trade menu tem s set to ON, the Range, Capacty, Graduaton, Zero Range, Auto Tare,
Retan Weght Data and Legal For Trade settngs are not reset.
.. Range
Set the number of weghng ranges.
SINGLE = one weghng range from zero to full capacty.
dUAL = two weghng ranges, where range 1 s from zero to half capacty and range 2 s from
half capacty to full capacty.
.. Capacity
Set the scale capacty as explaned n Secton 3.2 Menu Navgaton.
NOTE
: If dUAL was selected n the rANGE menu tem, the Capacty settng defnes the Range 2 capacty. The
Range 1 capacty s automatcally defned as half of the Capacty settng. For example, f Capacty s set to
15, the Range 1 capacty becomes 7.5.
After the capacty s set, select the Prmary Unt.
kg = the prmary unt s klograms
lb = the prmary unt s pounds