EN-1
5000 Series Indicators
.. Linearity Calibration
Lnearty calbraton uses 3 calbraton ponts. The full calbraton pont s establshed wth a weght on the
scale. The md calbraton pont s establshed wth a weght equal to half of the full calbraton weght on
the scale. The zero calbraton pont s establshed wth no weght on the scale. The md calbraton ponts
cannot be altered by the user durng the calbraton procedure.
When LINEAr s dsplayed, press the
Yes
button to ntate Lnearty Calbraton.
The dsplay flashes the full calbraton pont and calbraton unt. Place the specfed weght on the scale and
press the
Yes
button.
To choose a dfferent full pont or calbraton unt (kg or lb), edt the settng as explaned n Secton 3.2 Menu
Navgaton. When the desred settng s dsplayed, place the specfed weght on the scale and press the
Yes
button.
The dsplay shows --C-- whle the full pont s establshed.
The dsplay flashes the md calbraton pont.
Place the specfed weght on the scale and press the
Yes
button.
The dsplay shows --C-- whle the md pont s establshed.
The dsplay flashes 0.
Wth no weght on the scale, press the
Yes
button to establsh the zero pont.
The dsplay shows --C-- whle the zero pont s establshed.
When lnearty calbraton s completed, the dsplay shows dONE.
Then the scale exts to the actve weghng mode and dsplays the actual weght value.