▌For Calibration, press CALIBRATE Key.
Previous value
VFD Display Description
will be shown
▌Input current value shown on DVM Meter using cursor key and encoder
switch.
a-low 0.0603a
For example, if the value is 154.1mA, input as below.
VFD Display Description
▌When it is done, press CALIBRATE Key to save the data.
VFD Display Description
a-LOW 0.1541a
ADC-LO 00001B
DONE
▌To select HIGH range of current, press right cursor key once.
VFD Display Description
▌For Calibration, press CALIBRATE Key.
Previous value
VFD Display Description
will be shown
CAL - curr high
a-high 3.1078a
▌Input current value shown on DVM Meter using cursor key and encoder
switch.
For example, ife the value is 3.1235A, input as below.
VFD Display Description
▌When it is done, press CALIBRATE Key to save the data.
a-LOW 3.1235a
VFD Display Description
▌To apply in the system, power switch must be OFF and ON again.
▌Proceed other channel current as above instruction.
▌To cancel Calibration. Press Key.
aDC-HI 01FF1C
DONE
- 53 -