3-2. Constant Current Operating(CC)
It is CC output mode, and sequence are below :
▌Power Switch ON
After Power ON Check whether "P1 - OUTPUT OFF" is displayed
After Power ON, Check whether P1 - OUTPUT OFF is displayed.
▌Connect load with rear output terminal.
▌To set-up the limit, press LMT DISPLAY Key.
▌At first it is selected on the voltage, use cursor key and encoder switch to
change voltage value.
▌Press V/I Key once more to move cursor to current.
▌Use Cursor Key and Encoder Switch to set-up the current.
▌When all the set-up is done, press OUTPUT ON/OFF Key to allow
to print. It will automatically turn into ReadBack Display condition.
▌Check CC Lamp is ON, CV Lamp is OFF.
If it is ON and OFF opposite, check whether suppling voltage is enough,
and raise the voltage limit value.
≫ Related Remote Interface Command
INSTrument:[SELect] {OUT1|OUT2| .. |OUT255}
[SOURce:]VOLTage{<voltage>|UP|DOWN}
[SOURce:]CURRent{<current>|UP|DOWN}
OUTPut[:STATe] {OFF|ON|0|1}
OUTPut[:STATe] {OFF|ON|0|1}
Application:
OUTPUT OFF Output shut-off
INST OUT1
Channel P1 selected
VOLT 10
P1 Voltage 10V Setting
CURR 5
P1 Current 5A Setting
OUTPUT ON
Allow Voltage & Current Output
- 24 -