VOLTage:STEP?
Command which check setting Step Value.
Return Value "numeric value"
ex) volt:step?
return value "0.5000"
VOLTage:PROTection{<voltage>}
g
{
g
}
Command which set-up OVP Trip Level.
> Voltage Input voltage value in OVP setting range.
ex) volt:prot 32
OVP Level is 32V.
VOLTage:PROTection?
Command which checks OVP Trip Level.
Return Value "voltage"
g
ex)
return value "32.0000"
VOLTage:PROTection:STATe {0|1|OFF|ON}
Command which clear, ON OVP(Over Voltage Protection) .
> 0 OR OFF
OVP OFF
> 1 OR ON
OVP ON
ex) volt:prot:stat on
OVP On
VOLTage:PROTection:STAT?
Command which check whether OVP(Over Voltage Protection) is ON or cleared.
Return Value "0" - Cleared
"1" - ON
ex) volt:prot:stat?
return value "1"
VOLTage:PROTection:TRIPped?
Command which checks whether OVP(Over Voltage Protection) is occurred.
Return Value "0" - Normally activating
"1" - OVP Tripped
ex) volt:prot:trip?
return value "1"
Note
When OVP Trip occurs, output voltage/current will be shut-off.
VOLTage:PROTection:CLEar
Command which clears OVP(Over Voltage Protection) Trip.
Before clearing Trip please check "3-4. Programming Over Voltage Protection(OVP)" to find
g
p p
g
g
g
(
)
the reason of tripping
ex) volt:prot:cle
Clear OVP Trip.
- 67 -