≫ Related Remote Interface Command
INSTrument:[SELect] {OUT1|OUT2| .. |OUT255}
[SOURce:]CURRent:PROTection{<current>}
[SOURce:]CURRent:PROTection:STATe {0|1|OFF|ON}
Refer: CURR:PROT 6
OCP Level Setting
CURR:PROT:STAT ON or,
OCP ON
CURR:PROT:STAT OFF
OCP Cleared
CURR PROT STAT OFF
OCP Cleared
About the default OCP Setting, read reference "1-6. Check after Power ON" "Default Setting Value".
When actual output current is higher than OCP Level value, "OCP TRIPPED" will be displayed.
It is condition which shut-off all the source from relevant channel
Note
It is condition which shut-off all the source from relevant channel.
If it is in Tracking Mode or Parallel condition, please check.
To print voltage and current again, remove load first and clear trip please follow below sequenc
The cause of trips are :
First Mistaken OCP Level setting value may cause tripping. Therefore, raise Level value
Second Load which makes counter-electro motive force such as coils and motors may
Second Load which makes counter-electro motive force such as coils and motors may
Third When source one is used as load. When battery is run out, it may cause Trip.
▌
OCP TRIPPED
▌
When OCP trip occurs, "OCP TRIPPED" message will be displayed on VFD.
▌To clear OCP Trip, raise OCP Level first.
To set-up OCP level, Press OCP ->Cursor Key -> OCP Key.
VFD Display Description
▌
P
OCP K
i
"
OCP ON
"
diti
OCP ON
OCP - TRIPPED
▌
Press OCP Key once more in
"
OCP ON
" condition.
VFD Display Description
▌Use cursor key to select wanted digit.
▌Use encoder switch to raise the Level value.
▌Wh
i
i d
OCP K
OCP-LEVEL 3.20A
▌When setting is done, press OCP Key.
VFD Display Description
After displaying above message, it will return to previous page.
CHANGED
- 32 -