▌To guarantee precision, calibration should be done by a certificated organization every 6 months
and calibration once a year is necessary to use without problem. To avoid the worst case,
you can back-up calibration data of certificated organization.
▌Once back-up is done, the previous data can't be recovered.
CAL_BACKUP
▌Power on while pressing DYN SETTING,FACTORY key.
▌Power ON
▌Release the key after model name is displayed and entered
"--- FACTORY ---" mode.
▌Move to CAL_RESTORE menu.
"3.CAL_BACKUP" will be displayed.
FACTORY menu order: 1.USER_CLEAR ↔ 2.CAL_RESTORE ↔ 3.CAL_BACKUP ↔
4.CAL_DEFAULT ↔ 5.LOAD_DEFAULT ↔ 6.DELIMITER_CHAR ↔
7.COM_RESPONSE ↔ 8.AUTO_KEY_LOCK ↔ 9.ADC_SAMPLING
▌Press DYN SETTING,FACTORY key.
After "CAL.BACKUP_DONE" is displayed, the calibration value will be stored.
▌Power OFF
- 75 -
5-5. CAL_BACKUP
Press and hold
Release
DYN SETTING
FACTORY
POWER
DYN SETTING
FACTORY
POWER
DYN SETTING
FACTORY