Power Supply

▌Power supply can be used for source for calibration.
▌When you do current calibration, power supply's + terminal should be connected
   to load's + terminal  and load's - terminal should be connected to current monitoring
   shunt's one lead line. Other side lead line should be connected to power supply's 
   - terminal.

Current-Monitoring Resistor(shunt)

▌Use current monitoring resistor. Normal current measuring instrument is not suitable for
   high current capacity.
▌Select less than TCR 10ppm
▌Use 0.01% high precise standard resistor

DVM(Digital Volt Meter)

▌Used for voltage calibration and measuring current monitoring sensing voltage.
▌Resolution: 0.1 mV , Accuracy: more than 0.01%

Programming

This product supports calibration based on PC interface.
Calibration using communication doesn't cause measurement errors and
you can get accurate calibration data.
Below is the PC communication connection diagram.

< Figure 4-2 >

 -  62  - 

Summary of Contents for LF Series

Page 1: ...www odacore com Programmable DC Electronic Load LF Series User Manual Manual Part NO 018LF 1 0 1...

Page 2: ...d the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms the warranty terms in the separate agreement shall control Assist...

Page 3: ...en Make all connections to the unit before WARNING applying power Note the instrument s A WARNING notice denotes a hazard It calls wire firmly connected to an electrical ground A CAUTION notice denote...

Page 4: ...le push switch Out position of a bi stable push switch Caution risk of electric shock Earth ground terminal Protective earth ground terminal Frame or chassis terminal Terminal is at earth potential Ne...

Page 5: ...heck 1 4 Operating Conditions 3 1 5 Check Before Power On 4 Power Cable AC Input Power 1 6 Check After Power On 4 Displayed proceduer on LCD Default Setting Values 1 7 Installation 5 Cooling Bench Ope...

Page 6: ...nge LCD Display 3 5 ON OFF MODE 26 MODE SET Setting Value ON Current ON TIME OFF TIME ON OFF REPEAT Change LCD Display 3 6 FUSE MODE 29 MODE SET Setting Value FUSE Current FUSE TIME Change LCD Display...

Page 7: ...MOTE Voltage 58 3 15 SHORT 59 4 CALIBRATION 4 1 Features 60 4 2 Preparation for Calibration 60 4 3 Equipment required for Calibration 61 4 4 Measurement Techniques 61 Instruments Connection Drawing Po...

Page 8: ...74 5 5 CAL_BACKUP 75 5 6 CAL_DEFAULT 76 5 7 LOAD_DEFAULT 77 5 8 DELIMITER_CHAR 78 5 9 COM_RESPONSE 79 5 10 AOUT_KEY_LOCK 80 5 11 ADC_SAMPLING 81 6 1 Commands Syntax 82 6 2 Commands 82 7 1 Operating Er...

Page 9: ...7 4 Calibration Error 102 7 5 Nonvolatile Memory Check Error 103 7 6 Interface Commands Error 104 8 Cautions 106...

Page 10: ...nterface RS232C RS485 TCP IP Option SCPI Standard Commands for Programmable Instruments adopted High speed setting measument A large number of commands are programmed in product Easy interface setting...

Page 11: ...5M Output cable 1 pc Rack mount User s Manual 1 pc Options RS232C Cable 1M 2M 4M RS485 Cable 2 4 8 Channel 1M 2M 4M 10M AC Input Cable Special order type Output Cable Special order type 1 2 Accessorie...

Page 12: ...cal Check Check the broken key encoder switch power switch Check the panel surfaces are free of dents and scratches Check the broken rear output terminals Check the cabinet is free of scratches Check...

Page 13: ...n LCD All icons on LCD and segments will light up ODA Technologies is displayed on line 1 The model name of LF Series is displayed on line 2 During displaying above procedure the instrument is initial...

Page 14: ...ch Operation Load must be installed in a location that allows sufficient space at the sides front and rear of the unit for adequate air circulation There is no ventilation silt on the underside Rack M...

Page 15: ...1 CP MODE LCD ICON 28 DYNAMIC RUN STOP Calibration KEY 12 CR MODE LCD ICON 29 DYNAMIC SETING Factory KEY 13 CC MODE LCD ICON 30 INPUT Terminal 14 CV MODE LCD ICON 31 INPUT Terminal 15 Setting Encoder...

Page 16: ...N lights up when KEY LOCK mode In this mode you can t use any other keys 7 ERROR LCD ICON If errors occur the lamp lights up When all errors are checked the lamp goes out 8 V_SENSING ON LCD ICON light...

Page 17: ...CC CV 22 IO LOCAL KEY As Remote Interface setting key you can select RS232C RS485 and TCP IP In Remote Interface mode you can change to Local Mode using this key 23 ESC ERROR PROTECTION KEY Can be us...

Page 18: ...e During booting this key is for setting in Calibration menu and functional menu 29 DYNAMIC SETTING Factory KEY After booting the product this key is used to enter DYNAMIC MODE SETING menu During boot...

Page 19: ...ngeable 50 60Hz is default and change the value is optional Before power on be sure to check AC switch on the bottom Default is AC220V 2 External Input Output Option Terminal 1 V_Sensing Input 2 V_Sen...

Page 20: ...devices power supplier DMM Scope etc in parallel to one external controller Pin Description 1 5V Communication Power Source 3 D Data 2 D Data 4 0 Communication Ground 5 RS232C Interface Port This is a...

Page 21: ...t the operation of constant power mode 3 5 ON OFF MODE Description about ON OFF Mode of constant current 3 6 FUSE MODE Description about FUSE Mode of constant current 3 7 BAT MODE BATTERY Description...

Page 22: ...rors and disabling Protection 3 13 SAVE RECALL KEY LOCK Description about saving and loading the data from User Memory and disabling KEY LOCK mode 3 14 REMOTR Voltage Description about Remote Voltage...

Page 23: ...AVE is displayed on the 2nd line of LCD present voltage and current will displayed on the 1st line and I SET will displayed on the 2nd line of LCD Related Remote Interface Command MODE CC CV CR CP ONO...

Page 24: ...DISPLAY KEY and you can check various information when CC MODE is on DISPLAY change I SET WATT MODE TIME I SET When CC mode is on how to change current Check I SET setting When CC ON icon are on and...

Page 25: ...NPUT OFF To disable CC MODE press INPUT ON OFF KEY If INPUT is off ON icon will light off Related Remote Interface Command INPUT OFF ON INPUT Note Day Hour Minute Second MD T Unit 000 00 00 00 16 INPU...

Page 26: ...er SAVE is displayed on the 2nd line of LCD present voltage and current shows on the 1st line of LCD V SET will show on the 2nd line of LCD Related Remote Interface Command MODE CC CV CR CP ONOFF FUSE...

Page 27: ...AY KEY and you can check various information when CV MODE is on DISPLAY change V SET WATT MODE TIME V SET When CV mode is on how to change voltage Check V SET setting When CV and ON icon are on and V...

Page 28: ...NPUT OFF To disable CV MODE press INPUT ON OFF KEY If INPUT is off ON icon will light off Related Remote Interface Command INPUT OFF ON INPUT Note 19 Hour 000 00 00 00 Minute Second MD T unit Day INPU...

Page 29: ...fter SAVE is displayed on the 2nd line of LCD present voltage and current shows on the 1st line R SET will show on the 2nd line of LCD Related Remote Interface Command MODE CC CV CR CP ONOFF FUSE BAT...

Page 30: ...n DISPLAY change R SET WATT MODE TIME R SET When CR mode is on how to change resistance Check R SET setting When CR ON icon light on and R SET is displayed on the 2nd line LCD To set resistance move c...

Page 31: ...PUT OFF To disable CR MODE press INPUT ON OFF KEY If INPUT is off ON icon will light off Related Remote Interface Command INPUT OFF ON INPUT Note Day Hour Minute Second 00 00 00 000 MD TIME Unit 22 IN...

Page 32: ...u After SAVE is displayed on the 2nd line of LCD present voltage and current shows on the 1st line P SET will show on the 2nd line of LCD Related Remote Interface Command MODE CC CV CR CP ONOFF FUSE B...

Page 33: ...P MODE is on DISPLAY change P SET MODE TIME P SET When CP mode is on how to change watt Check P SET setting When CP and ON icon are on and P SET is displayed on the 2nd line LCD To set resistance move...

Page 34: ...PUT OFF To disable CP MODE press INPUT ON OFF KEY If INPUT is off ON icon will light off Related Remote Interface Command INPUT OFF ON INPUT Note MD TIME Unit 000 00 00 00 Second Hour 25 Day Minute IN...

Page 35: ...UE menu MODE ON OFF is displayed on 1st line of LCD and ON VALUE is displayed on 2nd line of LCD To set ON_VALUE move cursor using right left key to the value you want to change And adjust the encorde...

Page 36: ...d on 2nd line of LCD To set ON OFF_REP move cursor using right left key to the value you want to change And adjust the encorder switch to change value Saving MODE Exit MODE menu Press MODE key to save...

Page 37: ...ON_VALUE WATT ON OFF TIME MODE TIME REPEAT ON_VALUE ON OFF MODE INPUT OFF To disable ON OFF MODE OFF press INPUT ON OFF KEY if INPUT is off ON icon will light off Related Remote Interface Command INPU...

Page 38: ...MODE FUSE is displayed on 1st line of LCD VALUE is displayed on 2nd line of LCD To set FUSE value move cursor using right left key to the value you want to change And adjust the encorder switch to ch...

Page 39: ...CC 1 CV 2 CR 3 CP 4 ON OFF 5 FUSE 6 BAT FUSE CURR VALUE FUSE CURR FUSE TIME VALUE FUSE TIME FUSE MODE INPUT ON To enable ON OFF MODE on press INPUT ON OFF KEY If INPUT is on ON icon will light up Rela...

Page 40: ...is finished FUSE_MODE_END will be displayed on FUSE VALUE display and FUSE TIME will be stop and current value will be set as 0 When stop FUSE FUSE_CUT_OFF will be displayed on FUSE VALUE display and...

Page 41: ...T MODE menu MODE BAT is displayed on 1st line of LCD and BAT_MODE is displayed on 2nd line of LCD To set BAT_MODE CC CR CCP move cursor using right left key to the value you want to change BAT_MODE ch...

Page 42: ...LCD To set END CAPACITOR value move cursor using right left key to the value you want to change And adjust the encorder switch to change value Saving MODE Exit MODE menu Press MODE key to save MODE an...

Page 43: ...WATT BT_CAP BAT_VALUE BATTERY MODE INPUT OFF To disable ON OFF MODE off press INPUT ON OFF KEY If INPUT is off ON icon will light off Related Remote Interface Command INPUT OFF ON INPUT All values in...

Page 44: ...tep 1 FINISH_STEP is displayed on 2nd line of LCD Cycling setting munu order 1 FINISH_STEP 2 REPEAT_NUM 3 MODE 4 VALUE STEP 5 DELAY STEP Entering Finish Step menu Press Cycling setting key to enter Fi...

Page 45: ...eturn to Cycling setting menu Related Remote Interface Command CYC REP VALUE CYC REP Cycling MODE SET When in Cycling mode setting menu CYC icon is on and On icon is off Cycling Mode menu setting Move...

Page 46: ...e by using encorder switch Entering Cycling STEP value menu Press Cycling setting key once to enter Cycling STEP VALUE menu To set Cycling STEP value move cursor using right left key to the value you...

Page 47: ...EP DELAY menu To set Cycling STEP DELAY value move cursor using right left key to the value you want to change And adjust the encorder switch to change value 10mS unit Minimum 100mS Saving Cycling STE...

Page 48: ...STEP_VALUE Cycling mode STOP Check if Cycling is on CYC icon and On icon are light up To disable Cycling mode off press Cycling RUN STOP KEY CYC icon and On icon will light off and return to manual m...

Page 49: ...right key When in Dynamic menu 1 MODE is displayed on 2nd line of LCD Dynamic setting menu order 1 MODE 2 A VALUE 3 B VALUE CYC TIME 5 A DUTY Entering Dynamic mode menu Press Dynamic setting key once...

Page 50: ...r 1 MODE 2 A VALUE 3 B VALUE CYC TIME 5 A DUTY Dynamic A_VALUE menu Press Dynamic setting key once to enter Dynamic A_VALUE menu To set A VALUE move to REPEAT value which you want to set by using left...

Page 51: ...UE CYC TIME 5 A DUTY Entering Dynamic B Value menu Press Dynamic setting key once to enter Dynamic B Value menu To set B VALUE move cursor using right left key to the value you want to change And adju...

Page 52: ...UE CYC TIME 5 A DUTY Entering Dynamic CYC_TIME menu Press Dynamic setting key once to enter Dynamic CYC_TIME menu To set CYC TIME value move cursor using right left key to the value you want to change...

Page 53: ...mic A DUTY menu To set A DUTY value move cursor using right left key to the value you want to change And adjust the encorder switch to change value B DUTY is 100 A DUTY Saving A DUTY value Press Cycli...

Page 54: ...LAY change DYN A VALUE DYN B VALUE DYN CYC TIME A DUTY MD TIME DYN A VALUE Dynamic Mode STOP Check if Dynamic is on Dynamic icon and On icon are light up Dynamic mode STOP Press Dynamic RUN STOP KEY A...

Page 55: ...4 CC L_CV L Saving CC CV RANG Press CC CV RANG KEY After saving CC CV RANG exit CC CV RANG menu and return to manual mode Note CC CV RANG KEY is only available when INPUT of CC CV mode is OFF Setting...

Page 56: ...se check if the load is under KEY LOCK state RS 232C SET To set up RS 232C proceed as follows Power ON INPUT is off ON icon lights off Enter IO LOCAL menu Press IO Local Key to set RS 232C RS 232C men...

Page 57: ...wiring drawing in case both cables are Female type of standard Cross cable Refer 1 2 Accessories and Options to select properly by length and use Figure 3 2 If the remote device equipped DB25PIN only...

Page 58: ...nge RS 485 Baud Rate order BR_1 9600BPS BR_2 19200BPS BR_3 38400BPS BR_4 57600BPS BR_5 115200BPS Enter RS 485 ADDRESS menu Press IO Local key to set RS 485 ADDRESS Set RS 485 ADDRESS by using the enco...

Page 59: ...ram of Remote device PC by ODA s 485 to 232 converter Figure 3 5 RS485 PC Interface Drawing RS485 of LF Series outputs DC5V 0 2A It doesn t need any extra power supply because this model is designed t...

Page 60: ...der 1 RS 232C 2 RS 485 3 TCP IP Enter TCP IP menu Press IO Local Key to enter TCP IP menu Static IP setting Move to Static IP mode by using left right cursor When in Static IP mode TCP MODE_1 ST is di...

Page 61: ...mode DHCP SET Power ON INPUT is off ON icon is off Enter IO LOCAL menu Press IO Local Key to set TCP IP TCP IP setting Move to TCP IP menu by using left right key When in TCP IP menu 3 TCP IP is displ...

Page 62: ...nt Press IO Local Key to have TCP IP IP wait is displayed on the LCD Wait will be displayed while IP is being assigned After IP was assigned assigned IP will be displayed and Menu will be exited If IP...

Page 63: ...ucture type From 11th error errors will be discarded in the order they occured After check the message pressing ERROR Key it will be deleted in the order errors occurred When ERROR occurs alarm and er...

Page 64: ...on If heatsink s temperature is more than 100 Protection will be on and OVER TEMP PORT will be displayed To disable OTP stop operating load and cool down the load To disable OTP press ESC ERR PROT KEY...

Page 65: ...displayed on 2nd line SAVE RECALL menu order 1 SAVE MODE 2 RECAL MODE Enter SAVE mode menu Press SAVE RECALL V_SENSING KEY to enter SAVE Mode Menu Set SAVE address number Select address number between...

Page 66: ...MODE is displayed on 2nd line SAVE RECALL menu order 1 SAVE MODE 2 RECAL MODE Enter RECALL mode menu Press SAVE RECALL V_SENSING KEY to enter RECALL Mode Menu Setting RECALL address number Select addr...

Page 67: ...e output terminal to the load Please be careful of polarity Figure 3 6 Remote Voltage Mode Setting Power on V_SENSING OFF When INPUT is off ON SYN icons are off To enable V_SENSING ON press SAVE RECAL...

Page 68: ...ON icon lights up Press INPUT SHORT KEY LOCK KEY to enable SHORT MODE When SHORT ON SHORT ON is displayed on 2nd line Press INPUT SHORT KEY LOCK KEY to disable SHORT MODE off Then it return to CC MOD...

Page 69: ...input 60 4 1 Characteristic 4 CALIBRATION you need to calibrate your electronic load according to the environment Connect each measuring instruments without any contact resistance Turn off the electro...

Page 70: ...spec to make load s performace properly Measuring Instrument 4 3 Measuring instrument specifications Use Voltage Range 200 Vdc Current Range 60 Adc Based on LF 300A 61 Digital Voltmeter Voltage Calib...

Page 71: ...rrent monitoring resistor Normal current measuring instrument is not suitable for high current capacity Select less than TCR 10ppm Use 0 01 high precise standard resistor DVM Digital Volt Meter Used f...

Page 72: ...he front panel CALIBRATION KEY STRUCTURE 63 4 5 Calibration using front panel DYN RUN STOP CALIBRATE POWER ON 1 CC_HI_RNG DYN RUN STOP CALIBRATE 2 CC_LOW_RNG 3 CV_HI_RNG 4 CV_LOW_RNG CC_HI_LOW CC_HI_H...

Page 73: ...ne Power on Release the key after model name is displayed and entered CALIBRATION mode Move to 1 CC_HI_RNG and press DYNRUN STOP Calibration Key then CC_HI LOW will be displayed CALIBRATION menu order...

Page 74: ...ting press DYNRUN STOP Calibration key to save the setting value After HAX DAT is displayed go back to Calibration menu Move to 2 CC_LO_RNG and press DYNRUN STOP Calibration Key then CC_HI LOW will be...

Page 75: ...lue is 0 001 and the measured voltage is 6 12mV the current value is 6 12A Input the calculated current value to the load using cursor key and encoder switch After setting press DYNRUN STOP Calibratio...

Page 76: ...ey then CV_HI LOW will be displayed CALIBRATION menu order 1 CC_HI_RNG 2 CC_LOW_RNG 3 CV_HI_RNG 4 CV_LOW_RNG Wait until DVM s voltage becomes stable Input the calculated voltage value to the load usin...

Page 77: ...RUN STOP Calibration key to save the setting value After HAX DAT is displayed it will be changed to CV_LOW_HI_CAL mode CV_LOW HI will be displayed Wait until DVM s voltage becomes stable Input the cal...

Page 78: ...d to the load Command CAL CURR H Power supply s current value should be higher than load s max current value and set voltage as 5V If the power suppliy s current value is lower than load s calibration...

Page 79: ...delay is more than 500mS After several minutes measure DVM voltage connected to Current Monitoring resistance Send measured value to the power supply s voltage value For example if measured value is...

Page 80: ...et current several tens of mA If the voltage is less than or far greater than calibration range of the load CV can be measured incorrectly E g LF300 A CALIBRATION 155 160V 20 100mA After several minut...

Page 81: ...ta and set to default value 5 FACTORY 72 5 2 FACTORY KEY Strecture 5 1 Features DYN SETTING FACTORY POWER ON 1 USER_CLEAR DYN SETTING FACTORY 2 CAL_RESTORE 3 CAL_BACKUP 4 CAL_DEFAULT 5 LOAD_DEFAULT 6...

Page 82: ...and entered FACTORY mode Move to USER_CLEAR menu using cursor key 1 USER_CLEAR will be displayed FACTORY menu order 1 USER_CLEAR 2 CAL_RESTORE 3 CAL_BACKUP 4 CAL_DEFAULT 5 LOAD_DEFAULT 6 DELIMITER_CH...

Page 83: ...ntered FACTORY mode Move to CAL_RESTORE menu 2 CAL_RESTORE will be displayed FACTORY menu order 1 USER_CLEAR 2 CAL_RESTORE 3 CAL_BACKUP 4 CAL_DEFAULT 5 LOAD_DEFAULT 6 DELIMITER_CHAR 7 COM_RESPONSE 8 A...

Page 84: ...g DYN SETTING FACTORY key Power ON Release the key after model name is displayed and entered FACTORY mode Move to CAL_RESTORE menu 3 CAL_BACKUP will be displayed FACTORY menu order 1 USER_CLEAR 2 CAL_...

Page 85: ...RY key Power ON Release the key after model name is displayed and entered FACTORY mode Move to CAL_DEFAULT menu using cursor key 4 CAL_DEFAULT will be displayed FACTORY menu order 1 USER_CLEAR 2 CAL_R...

Page 86: ..._DEFAULT menu 5 LOAD_DEFAULT will be displayed FACTORY menu order 1 USER_CLEAR 2 CAL_RESTORE 3 CAL_BACKUP 4 CAL_DEFAULT 5 LOAD_DEFAULT 6 DELIMITER_CHAR 7 COM_RESPONSE 8 AUTO_KEY_LOCK 9 ADC_SAMPLING Pr...

Page 87: ...STORE 3 CAL_BACKUP 4 CAL_DEFAULT 5 LOAD_DEFAULT 6 DELIMITER_CHAR 7 COM_RESPONSE 8 AUTO_KEY_LOCK 9 ADC_SAMPLING Press DYN SETTING FACTORY key Enter DELIMITER_CHAR menu D CHAR will be displayed Move to...

Page 88: ...AL_DEFAULT 5 LOAD_DEFAULT 6 DELIMITER_CHAR 7 COM_RESPONSE 8 AUTO_KEY_LOCK 9 ADC_SAMPLING Press DYN SETTING FACTORY key Enter COM_RESPONSE menu RESP will be displayed Move to COM_RESPONSE using cursor...

Page 89: ...6 DELIMITER_CHAR 7 COM_RESPONSE 8 AUTO_KEY_LOCK 9 ADC_SAMPLING Press DYN SETTING FACTORY key Enter COM_RESPONSE menu A LOCK will be displayed on 2nd line Set AUTO_KEY_LOCK using cursor key AUTO_KEY_LO...

Page 90: ...LOAD_DEFAULT 6 DELIMITER_CHAR 7 COM_RESPONSE 8 AUTO_KEY_LOCK 9 ADC_SAMPLING Press DYN SETTING FACTORY key Enter COM_RESPONSE menu ADC_SAM will be displayed Set ADC_SAMPLING using cursor key ADC_SAMPL...

Page 91: ...han one One command in one time Square bracket means option or parameters and it can be omitted Do not omit the parameter in bracket The value in Triangular bracket can be changed by CODE ex MIN MAX B...

Page 92: ...ONOFF REP VALUE ONOFF REP ONOFF END FUSE CURR VALUE FUSE CURR FUSE TIME VALUE FUSE TIME FUSE END FUSE CUT BAT MODE CC CR CP BAT MODE BAT CURR VALUE BAT CURR BAT RES VALUE BAT RES BAT WATT VALUE BAT W...

Page 93: ...MEM A DYN MEM B VALUE DYN MEM B DYN FREQ VALUE DYN FREQ DYN DUTY VALUE DYN DUTY Cycling Commands CYC INPUT ON OFF CYC INPUT ON CYC MODE CC CV CYC MODE CYC VAL STEP_NUM VALUE CYC VAL STEP_NUM CYC DELA...

Page 94: ...Calibration Commands System Commands PROT PROT CLE SYST ERR IDN RST RCL VALUE SAV VALUE SN CAL VOLT MAX VALUE CAL CURR H L CAL CURR MIN VALUE CAL CURR MAX VALUE CAL VOLT H L CAL VOLT MIN VALUE 85...

Page 95: ...nt Voltage MODE CR CR Mode Constant Resitstance MODE CP CP Mode Constant Power MODE ON OFF ON OFF Mode Constant Current ON OFF test MODE FUSE FUSE Mode Constant Voltage test MODE BAT BAT Mode Battery...

Page 96: ...ltage ex VOLT Return value 10 000 RES value Command for setting input value of CR Mode Only available on CR Mode value Input resistance value ex RES 10 Set resistance as 10 ohms RES Command for checki...

Page 97: ...ME as 1S Make a space after ONOFF ON TIME It is Hour Minute Second Millisecond from the beginning in order without a space ONOFF ON TIME Command for checking ON Time of ON OFF Mode Return value time e...

Page 98: ...OFF Mode Return value 0 Before END Working 1 END Not working FUSE CURR VALUE Command for setting current value of FUSE Mode Only available on FUSE Mode value Input current value ex FUSE CURR 10 Set c...

Page 99: ...y available on BAT Mode BAT MODE CC CC Mode BAT MODE CR CR Mode BAT MODE CP CP Mode BAT MODE Command for checking mode of BAT Mode Return value 0 CC Mode 1 CR Mode 2 CP Mode BAT CURR VALUE Command for...

Page 100: ...value of BAT Mode Only available on BAT Mode value Input voltage ex BAT END VOLT 10 SET 10V BAT END VOLT Command for checking end voltage value of BAT Mode Return value voltage ex BAT END VOLT Return...

Page 101: ...rn value 1 0000 MEAS WATT Command for measuring the output watt of load Return value watt ex MEAS WATT return value 1 0000 Dynamic Commands Command for Dynamic Mode DYN INPUT ON OFF Command to enable...

Page 102: ...value 10 000 DYN MEM B VALUE Command for setting Bch HI value value Input Dynamic Bch value ex DYN MEM B 10 Set Dynamic as 10A or 10V Depending on selected Dynamic Mode CC CV voltage or current will...

Page 103: ...INPUT ON OFF Command to enable or disable Cycling Mode ON Enable Cycling Mode OFF Disable Cycling Mode ex1 CYC INPUT ON ex2 CYC INPUT OFF CYC INPUT Command for checking if Cycling Mode is on or off Re...

Page 104: ...rmation of Cycling Mode VALUE Cycling DELAY TIME ex CYC STEP DELAY 001 000000100 Save DELAY TIME 1S to Step 1 Set step 001 100 and make a space after CYC DELAY Input Hour 3digit Minute 2digit Second 2...

Page 105: ...10 Set step between 1 10 CYC SEQ Command for checking last step of Cycling Mode Return value final step ex CYC SEQ return value 10 CYC FIN Command for checking if repeat time of Cycling Mode is finish...

Page 106: ...ntering current High LOW Range Calibration set HI MAX current value VALUE ex CAL CURR MAX 60 123 CAL VOLT H L Command for entering voltage calibration H Command for HIGH RANGE voltage calibration L Co...

Page 107: ...s can be applied to serial number for distrubution when develop Windows application Return value LF 00 0000 00000 ex SN return value LF 03 0923 00185 IDN Command for checking load s information Three...

Page 108: ...latile User Memory 1 10 Memory save address ex sav 2 Save to address 2 RCL 1 2 3 4 5 6 7 8 10 Command for applying the data saved in non vilatile User Memory to your power supply Select one among 1 10...

Page 109: ...is not operating 202 Front panel operating failed Front panel is not operating 255 Error not define Error occured but not defined Refer to 4 6 REMOTE INTERFACE Calibration for GPIB for further informa...

Page 110: ...e Min VALUE MAX VALUE 28 Ignored min save under curr Occurs when operate MAX before operating Value of current MIN Procedure Min VALUE MAX VALUE 29 Invalid min value use under curr Occurs when operate...

Page 111: ...low limit over Occurs when over low range of Voltage ADC high Rang 64 DAC V low Rang high limit over Occurs when over high range of Voltage DAC low Rang 65 DAC V low Rang low limit over Occurs when ov...

Page 112: ...notify if an error occurred 80 Memory limit volt error There s an error in voltage limit value 81 Memory limit curr error There s an error in current limit value 82 Memory max volt error There s an e...

Page 113: ...vised volt 10 122 Syntax error There s a syntax error ex volt Value should be added after volt Revised volt 10 123 Invalid suffix There s an error in suffix of received data ex volt 10 should be remov...

Page 114: ...Revised volt 10 223 Incorret error Occurs when try a new operation without processing the buffer ex idn After send query command send a new command without acquiring data volt b data Revised idn a da...

Page 115: ...not place liquid containers on top of product liquid can cause serious damage to the product Don t apply strong vibration or pressure to product Acquire enough space near the ventilation slit Don t pu...

Page 116: ...62 Bupyeong daero 329beon gil Bupyeong gu Incheon city 403 858 Korea TEL 82 2 1800 8644 FAX 82 32 715 5456 www odacore com sales2 odacore com ODA TECHNOLOGIES CO LTD...

Reviews: