-23, "En route to cal the curr"
Occurs if you send calibration commands related to voltage during current calibration.
-24, "Over volt min parameter"
Occurs when over range of Voltage Min value.
Refer to "4-5. CALIBRATE '
-25, "Under volt max parameter"
Occurs when over lower limit value of Voltage Max.
Refer to "4-5. CALIBRATE '
-26, "Over volt max parameter"
Occurs when over upper limit value of Voltage Max.
Refer to "4-5. CALIBRATE '
-27, "Ignored min run under curr"
Occurs when operate MAX or VALUE before operating MiN value.
Procedure : Min → VALUE → MAX → VALUE
-28, "Ignored min save under curr"
Occurs when operate MAX before operating Value of current MIN.
Procedure : Min → VALUE → MAX → VALUE
-29, "Invalid min value use under curr"
Occurs when operate Min value again instead of MAX after operating current MIN value.
Procedure : Min → VALUE → MAX → VALUE
-30, "En route to cal the curr"
Occurs if you send calibration commands related to current during voltage calibration.
-31, "Over curr min parameter"
Occurs when over range of Current Min value.
Refer to "4-5. CALIBRATE"
-32, "Under curr max parameter"
Occurs when over lower limit value of Current Max.
Refer to "4-5. CALIBRATE"
-33, "Over curr max parameter"
Occurs when over upper limit value of Current Max.
Refer to "4-5. CALIBRATE"
- 101 -