▌After setting, press DYNRUN/STOP,Calibration key
to save the setting value.
After "HAX DAT: " is displayed, it will be changed to CC_LOW_HI_CAL mode.
"CC_LOW.HI" will be displayed.
▌Wait until DVM's voltage becomes stable.
▌After the voltage becomes stable, calculate the current value.
If the used resistance value is 0.001Ω and the measured voltage is 6.12mV,
the current value is 6.12A.
▌Input the calculated current value to the load
using cursor key and encoder switch.
▌After setting, press DYNRUN/STOP,Calibration key
to save the setting value.
After "HAX DAT: " is displayed, go back to Calibration menu
▌Power OFF
Note
When calibrating the current, lower the voltage of source by about 5V.
(Must use the power supply that can controll CV.)
Wait at least 3 mins
- 66 -