27
1568911, Edition 2019-04, Version 0
6
Mounting Threaded Bolts to the Interface Plate
EN
Figure 11: Cutting the threaded bolts for the Single / Duo inter
-
face plate on the raw ceiling to length
6.2 Mounting situation: Interface plate on the inter
-
mediate ceiling mount
6.2.1 Versions
(See "Figure 11")
The following Figures illustrate the interface plate
1
, Single (round
shape), for installing the motor arm and motor arm with extension arm
versions.
The steps required for installation on the interface plate
1
Single
(squared shape) are identical.
6.2.2 Cutting the threaded bolts to length
(See "Figure 11")
If a Single / Duo interface plate
1
is mounted to the raw ceiling with in
-
termediate ceiling
2
, the 6 (Single version) or 12 (Duo version) threaded
bolts M16 x 330mm① must be cut to length.
The Single / Duo canopy
3
to be mounted later is flush with the interme
-
diate ceiling
2
and covers the ceiling tube flange.
NOTE – Variable threaded bolt length
•
The necessary length of the threaded bolts M16 x 330mm①
depends on the distance H: from the raw ceiling to the bottom edge
of the intermediate ceiling
2
.
•
Observe the minimum and maximum length of the threaded bolts
M16 x 330mm①.
1. For a Single canopy
3
, 150mm, Duo canopy
3
, 150mm, cut the
6/12 threaded bolts M16 x 330mm ① to length.
•
To determine the length LG of the threaded bolts①:
LG=H +135mm (min. 205mm / max. 245mm)
•
Please contact Nuvo if you have any doubt.
2. Slightly deburr the 6/12 threaded bolts M16 x 330mm① in order to
ensure maximum thread engagement into the interface plate
1
.
3. Continue the installation as described in Chapter 6.2.3 on Page 28.