HOME PAGE LAYOUT
PAGE
|
84
Important Note:
On initial power-up of the laser, allow five seconds for the web interface to load
before accessing the web page.
When the home page opens, the I401 laser begins sending status data based on the laser’s current
condition. Once the initial data is loaded, the Get Data button becomes active. Click this button to begin
updating the static home page once a second. Click the Pause button to halt updating. If a fault should
occur while the home page is active, the updating process will automatically halt, and an error message
is displayed. Displayed data values are accurate to within ±0.25 units.
I401 web interface
Get Data button
Click to begin refreshing web page data once a second.
Pause button
Click to halt web page updates.
Status LEDs
Interlock, Temp, Ready, and Shutter icons display the current state of the laser’s Status LEDs.
Important Note: Because the i401 home page is a static web page, Ready and Shutter icons will
not flash fault codes; however, the gray Error messages section will display the fault.
RF module data
This section displays operating conditions for each of the i401’s four RF modules.
Connected
OK’ means that the RF module is connected to the Control module. ‘NC’ means the Control module does
not sense a connection to the RF module.
Input Voltage (VDC)
Displays the DC voltage level measured at the input of the Power module.
Switch Voltage (VDC)
Displays the DC voltage level measured at the input of the RF module.
Current (A)
Displays DC current (Amperage) being drawn by each RF module.
Temperature (°C)