MOUNTING
PAGE
|
21
surface. These holes should correspond with the fastener locations labeled “C” shown in the
figure below.
4
Carefully place the i401 laser on the mounting surface so the M10 threaded holes in the base plate
line up over the thru holes in the mounting surface.
5
Place a split washer and flat washer on each cap screw and insert the fasteners through the
mounting surface into the laser base plate. Turn the screws by hand until the threads fully
engage.
6
Evenly tighten all four fasteners to a torque of 40 N m (29 ft lb.).
Important Note: The i401 base plate contains two 6.40 mm × 12.70 mm (0.252" × 0.500") dowel
pin slots and a 6.40 mm (0.252") dowel pin hole for applications that require precision
positioning of the laser.
Figure 1-5 Mounting locations for four-point mount without feet.
Important Note: Verify the correct fastener length for your mounting application. The M10 × 1.5 ×
35 mm cap screws the ship kit are for use with the factory-installed mounting feet. When fastening
the i401 to your mounting surface from the bottom up, use M10 × 1.5 mounting screws with a length of
30 mm ± 2 mm plus the thickness of the mounting plate and any washers between the bolt head and the
mounting plate.