5.1 Preparations for Use
FRA5014
5-5
(Example 2)
•
Interface RS-232
•
Parity
Even
•
Stop bit
1
•
Baud rate
38400
Set SW1, SW5, and SW6 to “0” and SW2, SW3, SW4, SW7 and SW8 to “1”.
1
0
SW
1
2
3
4
5
6
7
8
5.1.5 Identification of USB devices
When the
FRA5014
is connected via a USB interface to a computer in which a USBTMC class
driver has been installed, the
FRA5014
is automatically recognized. The
FRA5014
in a system
can be identified according to the following parameters.
•
Vendor No. (Vendor ID):
3402 (0x0D4A)
Indicates NF Corporation
Expressed in decimal (hexadecimal) notation
•
Product number (Product ID):
18 (0x0012)
Indicates
FRA5014
•
Serial number (Serial):
Manufacturer’s number Unique number for each serial
number product unit
Serial number (manufacturer’s number) can be identified by the serial number label on the rear
panel.
5.1.6 GPIB address setting
With GPIB, an address specific to each device in the system is used to identify devices. Set a
unique GPIB address to each device. The
FRA5014
’s GPIB address is set via the rear panel’s DIP
switches.
Range of values: 0 to 30
(If an address of 31 or higher is set, operations assume it is 30.)