3.5 Basic Operations
FRA5014
3-12
3.5.2 Spot measurements
With spot measurements, one specified frequency point is measured, then its gain and phase are
measured for channel pairs CH2 & CH1, CH3 & CH1 and CH4 & CH1. Also, spot measurement
enables measurement result data to be read in complex formats.
The main spot measurement settings are listed below.
•
AC amplitude
•
DC bias
•
Measurement frequency
•
Measurement delay time
•
Integration time
•
Integration cycle
Setting example
*RST
Initializes
VOLT 1
Sets AC amplitude as 1 Vrms
VOLT:OFFS 0
Sets DC offset as 0 V
FREQ 10 kHz
Sets measurement frequency as 10 kHz
MEAS:DEL 0
Sets time between changing oscillation frequency and starting
measurement as 0 seconds
MEAS:INT:TIME 0.1
Sets integration time as 0.1 second
MEAS:INT:CYC 10
Sets integration cycles as 10 cycles
(The actual integration time (cycle) is the larger between the set integration time and the
integration cycle, and is set as an integer cycle period.)
VOLT:OUTP 2
Sets oscillator output as on
Measurement example
SWE:MEAS SPOT
Runs the spot measurement
SENS:DATA:SPOT?
Reads measurement data
(The MEASURE lamp is on during measurement. When the measurement is completed and
the MEASURE lamp goes off, read out the measurement data.)
After entering a remote command, set the unit to Talker.
For details, see
3.5.3 Measurement values during overload input
When the input signal to any of the analyzer’s channels exceeds the measurement range (±10 V), the
measured gain value is shown as the
FRA5014
’s maximum display value. Specifically, the
following measurement values are set.
•
Gain (dB)
+199.99 dB
•
Phase
0.00 deg
When overload input occurs in channel 1, the measurement values become as shown above. If
overload input occurs in channel 2, 3 or 4, the measurement values become as shown above only for
the channel with overload input.
Summary of Contents for FRA5014
Page 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Page 2: ......
Page 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Page 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Page 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Page 30: ......
Page 48: ......
Page 96: ......