7.8 Calibration
FRA5014
7-7
7.7.6 Dynamic range test
Connection:
OUTPUT OSCILLATOR
→
INPUT CH2, CH3, CH4. CH1 input is shorted (for CH1 test)
OUTPUT OSCILLATOR
→
INPUT CH1, CH3, CH4. CH2 input is shorted (for CH2 test)
OUTPUT OSCILLATOR
→
INPUT CH1, CH2, CH4. CH3 input is shorted (for CH3 test)
OUTPUT OSCILLATOR
→
INPUT CH1, CH2, CH3. CH4 input is shorted (for CH4 test)
Setting:
After initializing settings, set the frequency lower limit as 10 Hz, the frequency
upper limit as 100 kHz, frequency points as 40, integration time as 10 s, and AC
amplitude as 7.07 Vrms. Turn output on to have the settings reflected in the output.
Measurement:
Perform a sweep measurement and read the results.
If there is strong electromagnetic interference in the area, certain frequencies may
not be within their rated ranges. In such cases, remove the source of noise and try
again.
Judgment:
For CH1 test, normal if gain is > 120 dB; for CH2, CH3, and CH4 test, normal if
gain is <
−
120 dB.
7.8 Calibration
If the performance test shows that ratings are not met, contact NF Corporation to have adjustments
and/or calibration done in order to restore original performance.
When calibration is needed, contact NF Corporation or a sales representative.
A fee will be charged for adjustment or calibration performed after expiration of the warranty period.
Coaxial cable
CH1 dynamic range test example
CH1
CH2
OSCILLATOR
FRA
Input shorted
CH3
CH4
Summary of Contents for FRA5014
Page 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Page 2: ......
Page 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Page 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Page 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Page 30: ......
Page 48: ......
Page 96: ......