8.2 Analyzer Input Block
FRA5014
8-3
8.2 Analyzer Input Block
⚫
Number of channels
4
⚫
Input impedance
1 MΩ, 120 pF in parallel
⚫
Input voltage
Measurement range
±10 V
Non-destructive max. input ±24 V
⚫
Excessive input detection level
Detected by fundamental wave amplitude
Setting range
0.01 to 19.99 Vrms
Setting resolution
0.01 Vrms
Processing method
Beep sound output/sweep stop/oscillator off
⚫
Input weighting
For each channel,
Weighting coefficient: –1. 06 to +1. 06
Resolution: 6 digits or 0.01E
−
09
⚫
Measurement range
Auto switching
⚫
IMRR
At least 120 dB when the frequency is 1 Hz to 60 Hz and signal
source impedance
1 Ω
Attenuation when signal ground potential change (vs. case) is
mixed into the signal
⚫
Dynamic range
At least 120 dB
When the frequency is 1 Hz to 100 kHz, integration time
10 s
and
100 cycles
Signal magnitude ratio when 10 Vpeak signals are input to 3 of 4
channels while the remaining channel is terminated at a signal
source impedance of up to 1 Ω.
⚫
Isolation
Insulation voltage
42 Vpk or 30 Vrms, DC to 100 kHz continuous
Among each channel input in the analyzer block (signal, GND),
oscillator output (signal, GND), and the case
Case-to-oscillator
electrostatic capacitance
300 pF or less
Summary of Contents for FRA5014
Page 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Page 2: ......
Page 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Page 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Page 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Page 30: ......
Page 48: ......
Page 96: ......