7.4 Checking the Version Number
FRA5014
7-3
7.4 Checking the Version Number
The
FRA5014
’s firmware version number can be read by entering the query “*IDN?”.
For description of query commands,
7.5 Checking Isolation
The
FRA5014
’s oscillator output and analyzer input should each be insulated from the case.
Using a resistance meter (tester, multimeter, etc.), measure the resistance between the external
conductors on the rear panel’s BNC connectors and the grounding terminal in the center of the rear
panel, and make sure it is at least 10 MΩ. Similarly, check insulation between the BNC connectors.
7.6
Checking Oscillator’s Output Waveform
Connection:
OUTPUT OSCILLATOR
→
oscilloscope input (input impedance = 1 MΩ)
Setting:
After initializing settings, enter the remote commands “FREQ 1 kHz” and “VOLT
7.07” to set the frequency as 1 kHz and the amplitude as 7.07 Vrms, then enter the
remote command “VOLT:OUTP 2” to turn on the oscillator’s output.
Measurement:
Use an oscilloscope to observe the waveform.
Judgment:
Normal if 20 Vp-p sine wave. Make sure there is no distortion or noise.
7.7 Performance Test
This performance test is part of preventive maintenance aimed at preventing degradation of
performance in the
FRA5014
.
The performance test should be implemented during the
FRA5014
’s acceptance inspection, during
regular inspections, and during performance checks following repairs.
If the performance test results do not satisfy the rated specifications, calibration or repair is needed.
Contact NF Corporation or our representative.
Set the following before implementing a performance test.
•
Power supply voltage
100 to 230 V
10 %, but 250 V max.
•
Ambient temperature
23
5 °C
•
Ambient humidity
20 to 70 % RH, no condensation
•
Warm-up period
30 minutes or more
Note the following caution points before measuring performance.
•
If use of a coaxial cable is specified, use a cable with characteristic impedance of 50 Ω, thickness
of RG-58A/U or above, and length of one meter or less. The coaxial cable should have BNC
connectors at both ends.
•
When 50 Ω termination is specified, set 50 Ω for the input to the target measuring instrument. If
the measuring instrument cannot be set to 50 Ω, terminate by attaching a 50 Ω coaxial
feed-through terminator to the input of the measuring instrument.
•
The settings for the various test items are items to be changed when output is turned on after the
settings are initialized. Initialize all settings before implementing the test.
Summary of Contents for FRA5014
Page 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Page 2: ......
Page 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Page 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Page 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Page 30: ......
Page 48: ......
Page 96: ......