5.4 Command Descriptions
FRA5014
5-25
[:SOURce]:VOLTage:OFFSet[:IMMediate] <dc_offset>
[:SOURce]:VOLTage:OFFSet[:IMMediate]?
Description
Sets or queries the oscillator’s DC bias.
Parameter
<dc_offset> DC bias, format <NR2>, range
−
10.00 to 10.00 [unit: V]
Setting example
VOLT:OFFS 10.00
Response
<dc_offset>
Query example
VOLT:OFFS?
Response example 10.00
[:SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <amplitude> [VRMS|VPK]
[:SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Description
This sets or queries the oscillator’s AC amplitude.
This setting is interpreted as having the unit value set by the VOLTage:UNIT
setting.
However, if a numerical value includes a suffix (VRMS or VPK), it is interpreted
according to the suffix regardless of the VOLTage:UNIT setting.
Parameter
<amplitude> AC amplitude: <NR2> 0.000 to 7.07 [unit: Vrms]
0.000 to 10.00 [unit: Vpk]
Setting example
VOLT 7.07
Other example) VOLT 10 VPK, VOLT 1 VRMS
Response
<amplitude>
Query example
VOLT?
Response example 7.07
[:SOURce]:VOLTage:UNIT <amplitude_unit>
[:SOURce]:VOLTage:UNIT?
Description
Sets or queries the oscillator’s AC amplitude display unit and reponse unit.
Parameter
<amplitude_unit> AC amplitude unit, format <CRD>, range VRMS | VPK
Setting example
VOLT:UNIT VRMS
Response
<amplitude_unit>
Query example
VOLT:UNIT?
Response example VRMS
Summary of Contents for FRA5014
Page 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Page 2: ......
Page 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Page 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Page 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Page 30: ......
Page 48: ......
Page 96: ......