6.1 Error Displays
FRA5014
6-3
6.1.2 Remote control errors
Error
No.
Error message
Cause(s) of error
0 “No error”
No error
<Syntax errors>
−
100 “Command error”
An invalid command was received (no specific type).
−
102 “Syntax error”
An undefined command or parameter was received.
−
110 “Command header error”
Command header was recognized as invalid.
−
111 “Header separator error”
Header separator was not detected when parsing header.
<Execution-related errors>
−
200 “Execution error”
An error occurred when executing a command (no
specific type).
−
222 “Data out of range”
Command’s parameter is outside of setting range.
−
340 “Calibration failed”
Calibration failed.
−
350 “Queue overflow”
A newly occurred error cannot be retained due to error
queue overflow.
−
370 “Invalid (max<=min)”
Settings enabled as upper limit and lower limit settings
were reversed.
−
371 “OSC AC+DC > +/
−
10.5 V”
Oscillator’s output voltage peak is too high, so normal
signal output is not possible.
−
372 “OSC ac output = off”
Attempted to measure during zero AC output
−
373 “OSC is on”
Oscillator is ON; cannot execute self calibration
−
377 “In Calibrating”
Attempted to measure during calibration
−
378 “In Measuring”
Attempted calibration, etc., during measurement
−
381 “CH1 Overload”
Overload occurred in CH1
−
382 “CH2 Overload”
Overload occurred in CH2
−
383 “CH3 Overload”
Overload occurred in CH3
−
384 “CH4 Overload”
Overload occurred in CH4
* The error queue has four stages. If the error queue becomes full and another error occurs, error
350 (“Queue overflow”) is inserted at the end of the queue.
Summary of Contents for FRA5014
Page 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Page 2: ......
Page 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Page 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Page 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Page 30: ......
Page 48: ......
Page 96: ......