5.4 Command Descriptions
FRA5014
5-21
:SENSe:DATA:SPOT:LIMit:PHASe:MINimum <phase_lower_limit_ch2>,
<phase_lower_limit_ch3>, <phase_lower_limit_ch4>
:SENSe:DATA:SPOT:LIMit:PHASe:MINimum?
Description
Sets or queries the phase judgment lower limit for each channel’s spot
measurements.
1st parameter
<phase_lower_limit_ch2> CH2/CH1 phase judgment lower limit Range
−
180.00 to 180.00
2nd parameter
<phase_lower_limit_ch3> CH3/CH1 phase judgment lower limit Range
−
180.00 to 180.00
3rd parameter
<phase_lower_limit_ch4> CH4/CH1 phase judgment lower limit Range
−
180.00 to 180.00
Format is all <NR2> [unit: deg]
Setting example
SENS:DATA:SPOT:LIM:PHAS:MIN
−
120.00,
−
130.00,
−
100.00
Response
<phase_lower_limit_ch2>,<phase_lower_limit_ch3>,
<phase_lower_limit_ch4>
Query example
SENS:DATA:SPOT:LIM:PHAS:MIN?
Response example
−
120.00,
−
130.00,
−
100.00
:SENSe:DATA:SPOT:LIMit:REPort[:ALL]?
Description
Queries the spot judgment results. (All items are output.)
Parameters
None
Response 1
CH2/CH1 gain judgment results
−
1: NG if below lower limit / 0: SUCCESS / 1: NG if above upper limit
Response 2
CH2/CH1 phase judgment results
−
1: NG if below lower limit / 0: SUCCESS / 1: NG if above upper limit
Response 3
CH3/CH1 gain judgment results
−
1: NG if below lower limit / 0: SUCCESS / 1: NG if above upper limit
Response 4
CH3/CH1 phase judgment results
−
1: NG if below lower limit / 0: SUCCESS / 1: NG if above upper limit
Response 5
CH4/CH1 gain judgment results
−
1: NG if below lower limit / 0: SUCCESS / 1: NG if above upper limit
Response 6
CH4/CH1 phase judgment results
−
1: NG if below lower limit / 0: SUCCESS / 1: NG if above upper limit
Format is all <NR1>
Query example
SENS:DATA:SPOT:LIM:REP?
Response example 0,0,1,0,
−
1,
−
1
:SENSe:DATA:SPOT:LIMit:REPort:SELected? <gain_ch2>, <phase_ch2>, <gain_ch3>,
<phase_ch3>, <gain_ch4>, <phase_ch4>
Description
Outputs only the specified items (up to 7 items) among the spot judgment results.
1st parameter
<gain_ch2>
CH2/CH1 gain judgment output setting
0: No output / 1: Output
2nd parameter
<phase_ch2> CH2/CH1 phase judgment output setting 0: No output / 1: Output
3rd parameter
<gain_ch3>
CH3/CH1 gain judgment output setting
0: No output / 1: Output
4th parameter
<phase_ch3> CH3/CH1 phase judgment output setting 0: No output / 1: Output
5th parameter
<gain_ch4>
CH4/CH1 gain judgment output setting
0: No output / 1: Output
Summary of Contents for FRA5014
Page 1: ...NF Corporation SERVO ANALYZER FRA5014 Instruction Manual...
Page 2: ......
Page 3: ...SERVO ANALYZER FRA5014 Instruction Manual DA00019832 004...
Page 10: ...Contents FRA5014 vi 8 7 General Items 8 6...
Page 26: ...2 2 Installation FRA5014 2 8 Figure 2 5 Removal of Feet...
Page 30: ......
Page 48: ......
Page 96: ......