7.7 Operation Inspection
BP4610
7-7
7.7 Operation
Inspection
Verification before operation inspection
Before performing operation inspection, check the following items.
•
Is the supply voltage within the rated range?
•
Is the ambient temperature within the range of 5 to 35
°
C?
•
Is the ambient humidity within the range of 5 to 85%RH? (with absolute humidity of 1 to
25 g/m
3
)
•
Is there any condensation?
Check at power-on
Check that on error is displayed at power-on.
Check that the UNCAL icon and battery icon are not blinking.
(If an error is displayed,
refer to “6.1.1 Errors at power-on”.
)
LCD screen check
Check that the characters, icons, and lines displayed on the LCD screen are not displaced, do
not flicker, and free of bleeding.
Conditions during operation mode-specific function check
When performing the following operation-mode specific function checks, first initialize
(execute RESET in the SYSTEM menu) the settings for each check to prevent incorrect
settings.
Also, connect a 6
Ω
resistance load to the output terminal of the BP4610.
Check during CV mode internal signal source operation
Observe the voltage monitor output after connecting it to an oscilloscope. (The voltage of
1/100th the output voltage is monitored.)
In this state, try changing the settings listed below a number of times and check whether the
BP4610 functions normally.
DC voltage, superimposed AC voltage, superimposed AC frequency, superimposed AC
waveform, output on/off, voltage limiters (when the voltage limiters are set to a smaller value
than the output voltage while the BP4610 is outputting superimposed AC voltage, clipping of
the voltage frequency will occur and overload will be displayed on the LCD screen of the
BP4610).
Check during CC mode internal signal source operation
After performing setting initialization, set the operation mode to CC-INT.
Connect the current monitor output to an oscilloscope and observe it (observation of voltage
(V) that is 1/10th of the output current (A)).
In this state, try changing the settings listed below a number of times and check whether the
BP4610 functions normally.
DC current, superimposed AC current, superimposed AC frequency, superimposed AC
waveform, output on/off, current limiters (when the current limiters are set to a smaller value
than the output current while the BP4610 is outputting superimposed AC current, clipping of
the current waveform will occur and overload will be displayed on the LCD screen of the
BP4610).
Summary of Contents for BP4610
Page 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual nbn Austria GmbH...
Page 2: ......
Page 3: ...DA00010493 001 BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual...
Page 4: ......
Page 16: ......
Page 17: ...BP4610 1 OVERVIEW 1 1 General 1 1 1 2 Features 1 1 1 3 Principle of Operation 1 3...
Page 18: ......
Page 24: ......
Page 38: ......
Page 40: ......
Page 48: ......
Page 134: ......
Page 162: ......
Page 164: ......
Page 167: ...6 1 Error Messages BP4610 6 3 Figure 6 1 Protective Function Related Error Display Examples...
Page 174: ......
Page 190: ......
Page 207: ...BP4610 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 1...
Page 208: ......
Page 212: ......
Page 214: ......
Page 216: ......
Page 217: ......