2.4 Connection to I/O terminals
BP4610
2-10
2.4.5
Connection to external signal input pin
To use signal input from external as the signal source, connect the external signal input terminal (EXT
SIGNAL IN) on the front panel to the external signal source. For the connection, use a coaxial cable
with a BNC connector.
The specifications of the external signal input terminal are as follows.
Frequency range:
DC to 200 kHz
Gain: 100
×
(100 V/1 V), in-phase (CV mode)
10
×
(10 A/1 V), in-phase (CC mode)
Gain accuracy:
±
5% (1 kHz)
Input impedance:
10 k
Ω
Non-destructive max.:
±
5 V
POWER
EXT SIGNAL IN
V
O
I
O
MONITOR OUT
O I
1/100
1V/10A
SEQUENCE
TOP
MENU
START
STOP
HOLD
LO
OUTPUT
CANCEL
EN
BIPOLAR DC PO
MEASURE
EXT SIGNAL IN terminal
Figure 2-9. EXT SIGNAL IN Terminal
Summary of Contents for BP4610
Page 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual nbn Austria GmbH...
Page 2: ......
Page 3: ...DA00010493 001 BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual...
Page 4: ......
Page 16: ......
Page 17: ...BP4610 1 OVERVIEW 1 1 General 1 1 1 2 Features 1 1 1 3 Principle of Operation 1 3...
Page 18: ......
Page 24: ......
Page 38: ......
Page 40: ......
Page 48: ......
Page 134: ......
Page 162: ......
Page 164: ......
Page 167: ...6 1 Error Messages BP4610 6 3 Figure 6 1 Protective Function Related Error Display Examples...
Page 174: ......
Page 190: ......
Page 207: ...BP4610 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 1...
Page 208: ......
Page 212: ......
Page 214: ......
Page 216: ......
Page 217: ......