7.1 Introduction
BP4610
7-1
7.1 Introduction
To ensure use of the BP4610 under the best possible conditions, it is important to perform
maintenance as follows.
•
Operation verification:
Check performed to verify whether the BP4610 operates correctly.
•
Performance test:
Check performed to verify whether the BP4610 satisfies its
ratings.
•
Adjustment & calibration:
If the BP4610 does not satisfy its ratings, its performance can be
restored through adjustments and calibration by NF Corporation.
•
Troubleshooting:
If the above actions fail to yield improvements, NF Corporation
will investigate the cause(s) of the fault and the fault location(s),
and perform the necessary repairs.
This instruction manual explains operation inspections and performance tests that can be easily
performed. For more detailed inspections, adjustments, calibration, and troubleshooting, contact
NF Corporation or an NF Corporation sales representative.
Operation inspections and performance tests require the following measuring equipment.
•
Oscilloscope
Frequency band: 50 MHz and higher
•
Frequency counter
Standard oscillator stability: Within 10 ppm
•
AC voltmeter
TrueRMS type, accuracy: Within 0.2%, Frequency band: 300
kHz and higher
Recommended equipment: Agilent-made 34401A
•
DC voltage meter
Accuracy: Within 0.1%
Recommended equipment: Agilent-made 34401A
•
Distortion meter
Full scale: 1% or less, Frequency band: 100 kHz and higher
•
Sine wave oscillator
Amplitude: Within 3 Vp-p, Frequency: 10 Hz to 100 kHz,
Distortion: Within 0.2%
•
Resistance load
6
Ω
/600 W
•
Non-inductive current shunt
10 m
Ω
、
20 A or higher
7.2 Daily
Maintenance
Use the BP4610 in a place that meets the installation conditions.
For the installation conditions,
refer to “2.2 Installation Environment”.
If the panel or case surface is soiled, wipe it with a soft cloth. In the case of resistant stains, wipe
with a cloth that has been imbibed with a neutral detergent and then wrung well.
Wiping the BP4610 using an organic solvent such as thinner or benzene, or a premoistened wipe,
may cause degradation in material quality, clouding, or flaking off of the paint, and must
therefore be avoided.
Summary of Contents for BP4610
Page 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual nbn Austria GmbH...
Page 2: ......
Page 3: ...DA00010493 001 BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual...
Page 4: ......
Page 16: ......
Page 17: ...BP4610 1 OVERVIEW 1 1 General 1 1 1 2 Features 1 1 1 3 Principle of Operation 1 3...
Page 18: ......
Page 24: ......
Page 38: ......
Page 40: ......
Page 48: ......
Page 134: ......
Page 162: ......
Page 164: ......
Page 167: ...6 1 Error Messages BP4610 6 3 Figure 6 1 Protective Function Related Error Display Examples...
Page 174: ......
Page 190: ......
Page 207: ...BP4610 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 1...
Page 208: ......
Page 212: ......
Page 214: ......
Page 216: ......
Page 217: ......