7.8 Performance Test
BP4610
7-12
into a current value by dividing it by the load resistance.
Settings: After performing setting initialization, set the operation mode according to the
following table.
Judgment: If the output voltage or output current of the BP4610 falls within the rated ranges
in the following table, the operation is judged to be normal.
Operation Mode Setting
Rated Range
CV-EXT
50 mVrms or lower
CC-EXT
8 mArms or lower
(48 mVrms or lower @6
Ω
load)
Output voltage limiter accuracy test
Load: No load
Measurement connection: Measure the voltage of the output terminal of the BP4610 with a
DC voltmeter.
Settings: After performing setting initialization, set the output voltage limiters and DC output
voltage according to the following table.
Judgment: If the output voltage of the BP4610 falls within the rated ranges in the following
table, the operation is judged to be normal. This is the output voltage when overload has
occurred. The output is switched off after approximately 10 seconds.
+ Voltage Limiter Setting
−
Voltage Limiter Setting
DC Output Voltage Setting
Rated Range
+112 V
−
12 V
+115 V
+111 to +113 V
+12 V
−
112 V
−
115 V
−
113 to
−
111 V
Output current limiter accuracy test
Load: Short via a current shunt resistor
Measurement connection: Measure the detection voltage of the current shunt resistor with a
DC voltmeter and convert the measured value to a DC value.
Settings: After setting initialization has been performed, set the + current limiter to +10 A and
the – current limiter to
−
10 A.
Set the output voltage limiter and DC output voltage according to the following table.
Judgment: If the output current of the BP4610 falls within the rated ranges in the following
table, the operation is judged to be normal. This is the output current when overload has
occurred. The output is switched off after approximately 10 seconds.
+ Voltage Limiter Setting
−
Voltage Limiter Setting
DC Output Voltage Setting
Rated Range
+52 V
−
72 V
+10 V
+9 to +11 A
+72 V
−
52 V
−
10 V
−
11 to
−
9 A
Amplitude accuracy test of CV mode internal DC signal source
Load: No load
Measurement connection: Measure the voltage of the output terminal of the BP4610 with a
DC voltmeter.
Settings: After setting initialization has been performed, set the output voltage limiters and
DC output voltage according to the following table.
Judgment: If the output voltage of the BP4610 falls within the rated ranges in the following
table, the operation is judged to be normal.
+ Voltage Limiter Setting
−
Voltage Limiter Setting DC Output Voltage Setting
Rated Range
+117 V
−
7 V
+115 V
+114.23 to +115.77 V
+7 V
−
117 V
−
115 V
−
115.77 to
−
114.23 V
Summary of Contents for BP4610
Page 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual nbn Austria GmbH...
Page 2: ......
Page 3: ...DA00010493 001 BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual...
Page 4: ......
Page 16: ......
Page 17: ...BP4610 1 OVERVIEW 1 1 General 1 1 1 2 Features 1 1 1 3 Principle of Operation 1 3...
Page 18: ......
Page 24: ......
Page 38: ......
Page 40: ......
Page 48: ......
Page 134: ......
Page 162: ......
Page 164: ......
Page 167: ...6 1 Error Messages BP4610 6 3 Figure 6 1 Protective Function Related Error Display Examples...
Page 174: ......
Page 190: ......
Page 207: ...BP4610 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 1...
Page 208: ......
Page 212: ......
Page 214: ......
Page 216: ......
Page 217: ......