5.5 Response to Interface Messages
BP4610
5-27
5.5
Response to Interface Messages
The BP4610 allows USB interface control via the interface messages listed in Table 5-10.
Table 5-10. Interface Messages
Interface Message
Response
IFC Initialize
USB
interface.
SDC
Clear USB I/O buffer, clear status, and clear RQS bit.
REN
Set BP4610 to remote status.
GTL
Set BP4610 to local status.
LLO
Set BP4610 to local lockout status.
Disable [LOCAL] key on operation panel.
Perform lock so that switching to local mode via the operation panel is not performed.
5.6 Programming
Cautions
Arbitrary waveform data
The arbitrary waveform data is the binary data part, and therefore must be transmitted separately
from the command part of the preceding ASCII program message. The arbitrary waveform data must
be transferred as 1024 words (2048 bytes). If a different amount of data is transferred, an error will
occur or the correct waveform will not be output.
Example: 1) ARBT #42048
2) [higher byte of data 0][lower byte of data 0][higher byte of data 1]
[lower byte of data 1] …… [lower byte of data 1023]
Summary of Contents for BP4610
Page 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual nbn Austria GmbH...
Page 2: ......
Page 3: ...DA00010493 001 BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual...
Page 4: ......
Page 16: ......
Page 17: ...BP4610 1 OVERVIEW 1 1 General 1 1 1 2 Features 1 1 1 3 Principle of Operation 1 3...
Page 18: ......
Page 24: ......
Page 38: ......
Page 40: ......
Page 48: ......
Page 134: ......
Page 162: ......
Page 164: ......
Page 167: ...6 1 Error Messages BP4610 6 3 Figure 6 1 Protective Function Related Error Display Examples...
Page 174: ......
Page 190: ......
Page 207: ...BP4610 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 1...
Page 208: ......
Page 212: ......
Page 214: ......
Page 216: ......
Page 217: ......