6.1 Error Messages
BP4610
6-2
6.1.2
Protective function related errors
Table 6-2 lists the error messages displayed when protective function related errors occur.
Error messages are displayed in the message display window upon detection in the case of a power
failure, or the next time the BP4610 is powered on in all other cases. If an error message appears, press
the [ENTER] or [CANCEL] key.
(
Refer to “6.2 Protective Functions”.)
Table 6-2. Protective Function Related Errors
Error Message
Protection Item
Operation and Required Handling
Emergency shutdowned.
caused by:
OVERLOAD
Overload detection
anomaly
The message is displayed the next time the power is switched on.
Regardless of whether or not the output is off, the fact that
overload was detected is displayed.
Upon detection of this anomaly, contact NF Corporation or an
NF Corporation sales representative.
Emergency shutdowned.
caused by:
DCPS FAILURE
DCPS anomaly
The message is displayed the next time the power is switched on.
The message indicates that the fact that the DCPS output voltage
range exceeded the prescribed range was detected.
Upon detection of this anomaly, contact NF Corporation or an
NF Corporation sales representative.
Emergency shutdowned.
caused by:
OVER TEMPERATURE
Overheating anomaly
The message is displayed the next time the power is switched on.
This message indicates that overheating of the radiator was
detected.
Check if the air filter is clogged.
Emergency stop.
caused by:
POWER FAILURE
Power failure
The message is displayed upon detection of power failure.
It indicates that an instantaneous power interruption has been
detected.
Emergency shutdowned.
caused by:
BOOT FAILURE
Boot failure
The message is displayed the next time the power is switched on.
It indicates that the line monitoring circuit has not started up
normally.
Upon detection of this anomaly, contact NF Corporation or an
NF Corporation sales representative.
Emergency shutdowned.
caused by:
RESET FAILURE
Reset anomaly
The message is displayed the next time the power is switched on.
It indicates that the power-on reset circuit has not started up
normally.
Upon detection of this anomaly, contact NF Corporation or an
NF Corporation sales representative.
Emergency shutdowned.
caused by:
FRONT PANEL FAILURE
Operation panel
anomaly
The message is displayed the next time the power is switched on.
It indicates that an anomaly has been detected in the
communication data exchanged with the operation panel.
Upon detection of this anomaly, contact NF Corporation or an
NF Corporation sales representative.
Summary of Contents for BP4610
Page 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual nbn Austria GmbH...
Page 2: ......
Page 3: ...DA00010493 001 BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual...
Page 4: ......
Page 16: ......
Page 17: ...BP4610 1 OVERVIEW 1 1 General 1 1 1 2 Features 1 1 1 3 Principle of Operation 1 3...
Page 18: ......
Page 24: ......
Page 38: ......
Page 40: ......
Page 48: ......
Page 134: ......
Page 162: ......
Page 164: ......
Page 167: ...6 1 Error Messages BP4610 6 3 Figure 6 1 Protective Function Related Error Display Examples...
Page 174: ......
Page 190: ......
Page 207: ...BP4610 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 1...
Page 208: ......
Page 212: ......
Page 214: ......
Page 216: ......
Page 217: ......