7.8 Performance Test
BP4610
7-13
Amplitude accuracy test of CC mode internal DC signal source
Load: Short via a current shunt resistor
Measurement connection: Measure the detection voltage of the current shunt resistor with a
DC voltmeter and convert the measured value to a DC value.
Settings: After setting initialization has been performed, set the operation mode and DC
output current according to the following table.
Judgment: If the output current of the BP4610 falls within the rated ranges in the following
table, the operation is judged to be normal.
Operation Mode Setting
DC Output Current Setting
Rated Range
+10A
+9.92 to +10.08 A
CC-INT
−
10 A
−
10.08 to
−
9.92 A
Frequency accuracy test of internal AC signal source
Load: 6
Ω
resistance load
Measurement connection: Measure the signal (1 Vp-p) of the voltage monitor output terminal
of the BP4610 with a frequency counter.
Settings: After setting initialization has been performed, set the frequency and AC output
voltage according to the following table.
Judgment: If the output frequency of the voltage monitor of the BP4610 falls within the rated
range in the following table, the operation is judged to be normal.
Frequency Setting
AC Output Voltage Setting
Rated Range
10000 Hz
100 Vp-p
9999 to 10001 Hz
Amplitude accuracy test of CV mode internal AC signal source
Load: 6
Ω
resistance load
Measurement connection: Measure the voltage of the output terminal of the BP4610 with an
AC voltmeter.
Settings: After setting initialization has been performed, set the frequency and AC output
voltage according to the following table.
Judgment: If the output voltage of the BP4610 falls within the rated ranges in the following
table, the operation is judged to be normal.
Frequency Setting
AC Output Voltage Setting
Rated Range
120 Vp-p
119.2 to 120.8 Vp-p
(42.15 to 42.70 Vrms)
500 Hz
10 Vp-p
9.75 to 10.25 Vp-p
(3.446 to 3.623 Vrms)
Amplitude accuracy test of CC mode internal AC signal source
Load: Short via a current shunt resistor
Measurement connection: Measure the detection voltage of the current shunt resistor with an
AC voltmeter, and convert the measured value to a current value.
Settings: After setting initialization has been performed, set the operation mode, frequency,
and AC output current according to the following table.
Judgment: If the output current of the BP4610 falls within the rated ranges in the following
table, the operation is judged to be normal.
Operation Mode Setting
Frequency Setting
AC Output Current Setting
Rated Range
30 Ap-p
29.82 to 30.18 Ap-p
(10.55 to 10.67 Arms)
CC-INT 500
Hz
3 Ap-p
2.955 to 3.045 A
(1.045 to 1.076 Arms
)
Summary of Contents for BP4610
Page 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual nbn Austria GmbH...
Page 2: ......
Page 3: ...DA00010493 001 BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual...
Page 4: ......
Page 16: ......
Page 17: ...BP4610 1 OVERVIEW 1 1 General 1 1 1 2 Features 1 1 1 3 Principle of Operation 1 3...
Page 18: ......
Page 24: ......
Page 38: ......
Page 40: ......
Page 48: ......
Page 134: ......
Page 162: ......
Page 164: ......
Page 167: ...6 1 Error Messages BP4610 6 3 Figure 6 1 Protective Function Related Error Display Examples...
Page 174: ......
Page 190: ......
Page 207: ...BP4610 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 1...
Page 208: ......
Page 212: ......
Page 214: ......
Page 216: ......
Page 217: ......