4.10 Response Calibration
BP4610
4-78
(2) CC mode/inductive load
Basically, adjustments can be performed with the T and I adjustment knobs.
Turning the V adjustment knob to the right lowers the ringing frequency and makes it easier to
suppress ringing with the I adjustment knob, but normally the BP4610 is used with the V
adjustment knob turned all the way to the left.
The operations when the adjustment knobs are turned are described below.
•
If T adjustment knob is turned to the right
The rise/fall become faster, but the ringing frequency also rises, and its amplitude increases.
•
If I adjustment knob is turned to the right
The ringing amplitude is decreased but the rise/fall become slower.
To make the rise/fall as fast as possible and suppress ringing, perform adjustments by repeating
the following operations.
<1> Speed up the rise/fall with the T adjustment knob.
<2> Suppress ringing when it occurs with the I adjustment knob.
Table 4-25 shows adjustment examples for the three adjustment knobs, and Figures 4-26 to 4-28 show
output current waveform examples for each case.
Table 4-25. Adjustment Example for CC Mode/Inductive Load
Calibration Position
L
T V I
150
µ
H 5
1
7
390
µ
H 3
1
7
1.6 mH
1
1
7
Summary of Contents for BP4610
Page 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual nbn Austria GmbH...
Page 2: ......
Page 3: ...DA00010493 001 BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual...
Page 4: ......
Page 16: ......
Page 17: ...BP4610 1 OVERVIEW 1 1 General 1 1 1 2 Features 1 1 1 3 Principle of Operation 1 3...
Page 18: ......
Page 24: ......
Page 38: ......
Page 40: ......
Page 48: ......
Page 134: ......
Page 162: ......
Page 164: ......
Page 167: ...6 1 Error Messages BP4610 6 3 Figure 6 1 Protective Function Related Error Display Examples...
Page 174: ......
Page 190: ......
Page 207: ...BP4610 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 1...
Page 208: ......
Page 212: ......
Page 214: ......
Page 216: ......
Page 217: ......