3.127
EDH0162En1040 – 06/99
MM4005
TC — Read control loop type
Usage
IMM
PGM
MIP
Syntax
xxTC
Parameters
Description
xx
[int]
— Axis number.
Range
xx
—
1
to
4.
Units
xx
— None.
Defaults
xx
Missing: Error B.
Out of range: Error B.
Floating point: Error A.
Description
This command reads the type of control loop used on a specified axis. The
two possibilities are open loop (0) or closed loop (1).
NOTE
The controller can operate both stepper and DC motors in closed or open
loop. While the stepper motors operate fine in both modes, for normal
operation, DC motors should not be used in open loop.
Returns
xxTC nn
xx
—
Axis number.
nn
—
Loop type:
0 = Open loop.
1 = Closed loop.
Errors
A
—
Unknown message code.
B
—
Incorrect axis number.
S
—
Communication time-out.
Rel. Commands
SC
—
Set control loop type.
Example
1TC
|
Read the type of control loop used on axis #1.
1TC 1
|
Controller reports closed loop for axis #1.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com