3.23
EDH0162En1040 – 06/99
MM4005
AT — Tell the element number under execution
Usage
IMM
PGM
MIP
Syntax
AT
Parameters
None.
Description
This command retrieves from the controller the element number of the tra-
jectory that is currently being executed.
Returns
ATnn
nn
—
Element number.
Errors
S
—
Communication time-out.
Rel. Commands
XT
—
Tell number of elements in the trajectory.
LT
—
Extended list of the trajectory.
Example
ET |
Execute trajectory.
AT
|
Read current element number.
AT1
|
Controller returns 1.
AX — Assign a physical axis as X geometric axis
Usage
IMM
PGM
MIP
Syntax
xxAX
or
AX?
Parameters
Description
xx
[int]
— Physical axis number.
Range
xx
—
1
to
4.
Units
xx
— None.
Defaults
xx
Missing: Error B.
Out of range: Error B.
Description
This command tells to the controller which physical axis will be the X geo-
metric axis for the next trajectory that will be loaded. ET command will ver-
ify the correct assignation in execution.
Returns
If AX? takes place of the xxAX, this command reportes the actual number of
assigned as X geometric axis.
Errors
B
—
Incorrect axis number.
S
—
Communication time-out.
Rel. Commands
AY
—
Assign a physical axis as Y geometric axis.
LT
—
Extended list of the trajectory.
Example
1AX |
Assign physical axis 1 as X geometric axis.
AX? |
AX1
|
Controller return value 1.