New Brunswick Scientific Co., Inc.
User’s Guide
79
2 CH COMP
•
OUTPUT A
varies linearly from
minimum
(Complementary) at 100% controller output, to
maximum
at +100% controller output.
•
OUTPUT B
varies linearly from
maximum
at -100%
controller output to
minimum
at 100% controller output.
This action allows two actuators with a continuous transition
between them.
2 CH OVER
•
OUTPUT A
varies linearly from
minimum
(Overlap) at
-25% controller output, to
maximum
at 100% controller
output.
•
OUTPUT B
varies linearly from
maximum
at -100%
controller output to
minimum
at +25% controller output.
This action allows an overlap control transition between two
actuators.
DIRECT w/
ENABLE
Same controller action
DIRECT
,
with an additional digital
controller output which is
OFF
when either controller output
or setpoint is zero,
ON
when both setpoint and controller
output are not zero. This function is used to actuate the
shutdown features of mass flow controllers, motor
controllers, etc.
18. Select
ENABLE AUTO STERILIZE
only if this loop (typically, temperature) will
control the sterilization cycle.
19. If you wish to calibrate the control loop through the controller, select
ENABLE
CALIBRATION
. If this box is not checked, the control loop will not be visible
in the
SELECT CALIBRATION SCREEN
. Refer to the
Calibration
section of
this manual for detailed instructions on loop calibration. The BioFlo4500 is
configured at the factory to display the following loops on the
SELECT
CALIBRATION SCREEN
:
The factory settings for the select calibration screen are:
•
Pointing Device
•
Pump 1
•
Aux Pump
•
pH
•
Pump 2
•
Hi Foam
•
DO
•
Pump 3
•
Anti Foam
Summary of Contents for BioFlo 4500
Page 2: ...BioFlo 4500 M1254 0050 User s Guide ii...
Page 4: ...BioFlo 4500 M1254 0050 User s Guide iv THIS PAGE INTENTIONALLY BLANK...
Page 6: ...BioFlo 4500 M1254 0050 User s Guide vi THIS PAGE INTENTIONALLY BLANK...
Page 8: ...BioFlo 4500 M1254 0050 User s Guide viii THIS PAGE INTENTIONALLY BLANK...
Page 10: ...BioFlo 4500 M1254 0050 User s Guide x THIS PAGE INTENTIONALLY BLANK...
Page 12: ...BioFlo 4500 M1254 0050 User s Guide xii THIS PAGE INTENTIONALLY BLANK...
Page 18: ...BioFlo 4500 M1254 0050 User s Guide xviii...
Page 34: ...BioFlo 4500 M1254 0050 User s Guide 34 THIS PAGE INTENTIONALLY BLANK...
Page 42: ...BioFlo 4500 M1254 0050 User s Guide 42 THIS PAGE INTENTIONALLY BLANK...
Page 60: ...BioFlo 4500 M1254 0050 User s Guide 60 THIS PAGE INTENTIONALLY BLANK...
Page 82: ...BioFlo 4500 M1254 0050 User s Guide 82 THIS PAGE INTENTIONALLY BLANK...
Page 88: ...BioFlo 4500 M1254 0050 User s Guide 88 THIS PAGE INTENTIONALLY BLANK...
Page 114: ...BioFlo 4500 M1254 0050 User s Guide 114 THIS PAGE INTENTIONALLY BLANK...
Page 138: ...BioFlo 4500 M1254 0050 User s Guide 138 THIS PAGE INTENTIONALLY BLANK...
Page 144: ...BioFlo 4500 M1254 0050 User s Guide 144 THIS PAGE INTENTIONALLY BLANK...
Page 152: ...BioFlo 4500 M1254 0050 User s Guide 152 THIS PAGE INTENTIONALLY BLANK...
Page 154: ...BioFlo 4500 M1254 0050 User s Guide 154...
Page 160: ...BioFlo 4500 M1254 0050 User s Guide 160 THIS PAGE INTENTIONALLY BLANK...
Page 162: ...BioFlo 4500 M1254 0050 User s Guide 162 THIS PAGE INTENTIONALLY BLANK...
Page 164: ...BioFlo 4500 M1254 0050 User s Guide 164 THIS PAGE INTENTIONALLY BLANK...
Page 168: ...BioFlo 4500 M1254 0050 User s Guide 168 THIS PAGE INTENTIONALLY BLANK...
Page 186: ...BioFlo 4500 M1254 0050 User s Guide 186 THIS PAGE INTENTIONALLY BLANK...
Page 194: ...BioFlo 4500 M1254 0050 User s Guide 194 THIS PAGE INTENTIONALLY BLANK...
Page 196: ...BioFlo 4500 M1254 0050 User s Guide 196 THIS PAGE INTENTIONALLY BLANK...
Page 208: ...BioFlo 4500 M1254 0050 User s Guide 208 THIS PAGE INTENTIONALLY BLANK...
Page 214: ...BioFlo 4500 M1254 0050 User s Guide 214 Figure 49 Power Schematics Detail A...