2. Reduce the vessel pressure to 0 PSIG using the
BACKPRESSURE
REGULATOR
located in the upper left rear corner of the open piping
frame (manual pressure control), or set the setpoint to “O” in the
PRESSURE
DISPLAY GAUGE
on the touchscreen
.
3. Assuming the port was previously sterilized, carefully remove the port
steam cap and aseptically attach a harvest vessel using silicone or
Tygon type tubing.
4. To open the drain, rotate the drain valve handle as far as it will allow
so it is facing the front of the vessel. When the valve is open, the word
OPEN
is visible on the end of the handle.
5. Draining of the liquid may be accomplished by gravity, a peristaltic
pump, or by using the vessel pressure to drive out the liquid from
inside the vessel.
6. After draining the vessel, close the drain valve by rotating the drain
valve 180
°
. In the closed position, the handle should be facing the
front panel console and the word
CLOSE
is visible on the end the
handle.
7. Break the harvest vessel connection by removing the connecting
tubing from the quick-connect fitting.
8. Put the stainless steel port steam cap back on the port and re-sterilize
the port.
6.4.2
Asceptic Harvest/Drain Method B
1. Prepare an appropriate harvest vessel fitted with tubing and a quick-
connect fitting. Sterilize if appropriate.
NOTE:
Refer to your local NBS parts distributor or sales office for quick-
connect options.
2. Autoclave harvest vessel and quick-connect fitting in the closed
position (121ºC, 30 minutes minimum, depending on vessel).
3. Remove steam port cap from the harvest/drain port and connect the
quick-connect fitting from the harvest vessel to the harvest/drain port.
Summary of Contents for BioFlo 4500
Page 2: ...BioFlo 4500 M1254 0050 User s Guide ii...
Page 4: ...BioFlo 4500 M1254 0050 User s Guide iv THIS PAGE INTENTIONALLY BLANK...
Page 6: ...BioFlo 4500 M1254 0050 User s Guide vi THIS PAGE INTENTIONALLY BLANK...
Page 8: ...BioFlo 4500 M1254 0050 User s Guide viii THIS PAGE INTENTIONALLY BLANK...
Page 10: ...BioFlo 4500 M1254 0050 User s Guide x THIS PAGE INTENTIONALLY BLANK...
Page 12: ...BioFlo 4500 M1254 0050 User s Guide xii THIS PAGE INTENTIONALLY BLANK...
Page 18: ...BioFlo 4500 M1254 0050 User s Guide xviii...
Page 34: ...BioFlo 4500 M1254 0050 User s Guide 34 THIS PAGE INTENTIONALLY BLANK...
Page 42: ...BioFlo 4500 M1254 0050 User s Guide 42 THIS PAGE INTENTIONALLY BLANK...
Page 60: ...BioFlo 4500 M1254 0050 User s Guide 60 THIS PAGE INTENTIONALLY BLANK...
Page 82: ...BioFlo 4500 M1254 0050 User s Guide 82 THIS PAGE INTENTIONALLY BLANK...
Page 88: ...BioFlo 4500 M1254 0050 User s Guide 88 THIS PAGE INTENTIONALLY BLANK...
Page 114: ...BioFlo 4500 M1254 0050 User s Guide 114 THIS PAGE INTENTIONALLY BLANK...
Page 138: ...BioFlo 4500 M1254 0050 User s Guide 138 THIS PAGE INTENTIONALLY BLANK...
Page 144: ...BioFlo 4500 M1254 0050 User s Guide 144 THIS PAGE INTENTIONALLY BLANK...
Page 152: ...BioFlo 4500 M1254 0050 User s Guide 152 THIS PAGE INTENTIONALLY BLANK...
Page 154: ...BioFlo 4500 M1254 0050 User s Guide 154...
Page 160: ...BioFlo 4500 M1254 0050 User s Guide 160 THIS PAGE INTENTIONALLY BLANK...
Page 162: ...BioFlo 4500 M1254 0050 User s Guide 162 THIS PAGE INTENTIONALLY BLANK...
Page 164: ...BioFlo 4500 M1254 0050 User s Guide 164 THIS PAGE INTENTIONALLY BLANK...
Page 168: ...BioFlo 4500 M1254 0050 User s Guide 168 THIS PAGE INTENTIONALLY BLANK...
Page 186: ...BioFlo 4500 M1254 0050 User s Guide 186 THIS PAGE INTENTIONALLY BLANK...
Page 194: ...BioFlo 4500 M1254 0050 User s Guide 194 THIS PAGE INTENTIONALLY BLANK...
Page 196: ...BioFlo 4500 M1254 0050 User s Guide 196 THIS PAGE INTENTIONALLY BLANK...
Page 208: ...BioFlo 4500 M1254 0050 User s Guide 208 THIS PAGE INTENTIONALLY BLANK...
Page 214: ...BioFlo 4500 M1254 0050 User s Guide 214 Figure 49 Power Schematics Detail A...