BioFlo 4500 M1254-0050
User’s Guide
24
Ports
Description
Fitting/Size
Location
Notes
Spare
Ingold Type
25 mm
Lower Side Wall For insertion of redundant
probe or third party sensors
or samplers.
Redox or
turbidity probes available as
options: please inquire.
Harvest/Drain Valve
Plunger-type
NBS Valve
Bottom
Standard, equipped with 1/4"
quick-connect fitting
Jacket Pressure Relief
Valve
1/2" NPT
Jacket
Standard, rated at 35 PSIG
Jacket Water Out
1/2" NPT
Jacket
Standard
Jacket Water In
1/2" NPT
Jacket
Standard
Piping and Instrumentation with ML-6100 Controller
Materials Process
All process piping, valves, and associated components
are Type 316L stainless steel
Non-Process
Type 316L or Type 304 stainless steel
Gaskets
Silicone
O-rings
Ethylene Propylene EPDM
Temperature Sensor
Platinum RTD sensor in thermowell
Actuator
Four high watt density cartridge heaters in the bottom of
the vessel (675W in 20L vessel & 850W in 30L vessel).
Cooling water is admitted into the jacket through the
temperature control valve and the flow control valve.
Control
Mode
PID with constants factory-set at P = 39, I = 1.3,
D =0, PWM 0-100%, 2 Ch Direct, and Period 10
Range
5
°
C above coolant supply temperature to a maximum of
85
°
C
±
0.2
°
C
Display
Current value, setpoint, controller output % displayed
on virtual display gauge at touchscreen
Agitation Speed
Sensor
Frequency 1000 Line Encoder
Drive
1 HP, direct drive, AC, top-mounted motor. Allows OTR
of 350 mMoles/L/hr
Control
Mode
Direct w/ Enable and PID with constants factory-set at P
= 0.30, I = 4.0, D =0
Range
50-1000 rpm
±
1 rpm
Bearing
Housing
Standard single or optional double mechanical seals
Display
Current value, setpoint, controller output % displayed
on virtual display gauge at touchscreen
Aeration
Air inlet filter
Steam-sterilizable absolute 0.2µ filter that continuously
removes viable organisms, cartridge- type 316 L,
stainless steel housing, Domnick Hunter.
NOTE: Relative humidity must be less than 95%.
Other brands can be supplied upon request
…continued…
Summary of Contents for BioFlo 4500
Page 2: ...BioFlo 4500 M1254 0050 User s Guide ii...
Page 4: ...BioFlo 4500 M1254 0050 User s Guide iv THIS PAGE INTENTIONALLY BLANK...
Page 6: ...BioFlo 4500 M1254 0050 User s Guide vi THIS PAGE INTENTIONALLY BLANK...
Page 8: ...BioFlo 4500 M1254 0050 User s Guide viii THIS PAGE INTENTIONALLY BLANK...
Page 10: ...BioFlo 4500 M1254 0050 User s Guide x THIS PAGE INTENTIONALLY BLANK...
Page 12: ...BioFlo 4500 M1254 0050 User s Guide xii THIS PAGE INTENTIONALLY BLANK...
Page 18: ...BioFlo 4500 M1254 0050 User s Guide xviii...
Page 34: ...BioFlo 4500 M1254 0050 User s Guide 34 THIS PAGE INTENTIONALLY BLANK...
Page 42: ...BioFlo 4500 M1254 0050 User s Guide 42 THIS PAGE INTENTIONALLY BLANK...
Page 60: ...BioFlo 4500 M1254 0050 User s Guide 60 THIS PAGE INTENTIONALLY BLANK...
Page 82: ...BioFlo 4500 M1254 0050 User s Guide 82 THIS PAGE INTENTIONALLY BLANK...
Page 88: ...BioFlo 4500 M1254 0050 User s Guide 88 THIS PAGE INTENTIONALLY BLANK...
Page 114: ...BioFlo 4500 M1254 0050 User s Guide 114 THIS PAGE INTENTIONALLY BLANK...
Page 138: ...BioFlo 4500 M1254 0050 User s Guide 138 THIS PAGE INTENTIONALLY BLANK...
Page 144: ...BioFlo 4500 M1254 0050 User s Guide 144 THIS PAGE INTENTIONALLY BLANK...
Page 152: ...BioFlo 4500 M1254 0050 User s Guide 152 THIS PAGE INTENTIONALLY BLANK...
Page 154: ...BioFlo 4500 M1254 0050 User s Guide 154...
Page 160: ...BioFlo 4500 M1254 0050 User s Guide 160 THIS PAGE INTENTIONALLY BLANK...
Page 162: ...BioFlo 4500 M1254 0050 User s Guide 162 THIS PAGE INTENTIONALLY BLANK...
Page 164: ...BioFlo 4500 M1254 0050 User s Guide 164 THIS PAGE INTENTIONALLY BLANK...
Page 168: ...BioFlo 4500 M1254 0050 User s Guide 168 THIS PAGE INTENTIONALLY BLANK...
Page 186: ...BioFlo 4500 M1254 0050 User s Guide 186 THIS PAGE INTENTIONALLY BLANK...
Page 194: ...BioFlo 4500 M1254 0050 User s Guide 194 THIS PAGE INTENTIONALLY BLANK...
Page 196: ...BioFlo 4500 M1254 0050 User s Guide 196 THIS PAGE INTENTIONALLY BLANK...
Page 208: ...BioFlo 4500 M1254 0050 User s Guide 208 THIS PAGE INTENTIONALLY BLANK...
Page 214: ...BioFlo 4500 M1254 0050 User s Guide 214 Figure 49 Power Schematics Detail A...