BioFlo 4500 M1254-0050
User’s Guide
134
17.1.2
Sparging with Nitrogen to Calibrate DO Zero Point
A more accurate zero setting is obtained by assuring the probe has
stabilized at the desired growth temperature and then by setting agitation
speed to 500 rpm and sparging nitrogen into the vessel. After the DO
reading stabilizes for at least 5 minutes, enter the “zero” on the
PROBE
CALIBRATION SCREEN
.
To calibrate the span:
To calibrate the span value, the fermentor should be on and operating at
the desired growth temperature. Then proceed as follows:
•
Set the agitation speed to 500 rpm and vigorously sparge air into the
vessel.
•
Carefully monitor the current value reading of DO control loop display
gauge. Once the reading has stabilized, press the
CAL
BUTTON
. The
CALIBRATION SELECTION WINDOW
will appear.
•
Select the DO control loop from the list by pressing on it. The
LOOP
CAILBRATION SCREEN
will appear.
•
To set the span, ensure that the
SET SPAN FUNCTION
is selected, and
enter 100 into the span text box using the onscreen keyboard. Once
the number is entered, press the
SET SPAN
button. The value entered
will then become the span value.
•
Press
CLOSE
to exit this window.
NOTE:
The probe is calibrated in percentage air saturation at a given
temperature, a given pressure (calibration conditions). If pressure is
raised, then readings may be obtained which are higher than 100%.
NOTE:
Monitoring DO during calibration can be done by viewing the DO
gauge or by setting up a Trend Graph for DO and viewing the graph.
(See section on Creating a Trend Graph)
.
Summary of Contents for BioFlo 4500
Page 2: ...BioFlo 4500 M1254 0050 User s Guide ii...
Page 4: ...BioFlo 4500 M1254 0050 User s Guide iv THIS PAGE INTENTIONALLY BLANK...
Page 6: ...BioFlo 4500 M1254 0050 User s Guide vi THIS PAGE INTENTIONALLY BLANK...
Page 8: ...BioFlo 4500 M1254 0050 User s Guide viii THIS PAGE INTENTIONALLY BLANK...
Page 10: ...BioFlo 4500 M1254 0050 User s Guide x THIS PAGE INTENTIONALLY BLANK...
Page 12: ...BioFlo 4500 M1254 0050 User s Guide xii THIS PAGE INTENTIONALLY BLANK...
Page 18: ...BioFlo 4500 M1254 0050 User s Guide xviii...
Page 34: ...BioFlo 4500 M1254 0050 User s Guide 34 THIS PAGE INTENTIONALLY BLANK...
Page 42: ...BioFlo 4500 M1254 0050 User s Guide 42 THIS PAGE INTENTIONALLY BLANK...
Page 60: ...BioFlo 4500 M1254 0050 User s Guide 60 THIS PAGE INTENTIONALLY BLANK...
Page 82: ...BioFlo 4500 M1254 0050 User s Guide 82 THIS PAGE INTENTIONALLY BLANK...
Page 88: ...BioFlo 4500 M1254 0050 User s Guide 88 THIS PAGE INTENTIONALLY BLANK...
Page 114: ...BioFlo 4500 M1254 0050 User s Guide 114 THIS PAGE INTENTIONALLY BLANK...
Page 138: ...BioFlo 4500 M1254 0050 User s Guide 138 THIS PAGE INTENTIONALLY BLANK...
Page 144: ...BioFlo 4500 M1254 0050 User s Guide 144 THIS PAGE INTENTIONALLY BLANK...
Page 152: ...BioFlo 4500 M1254 0050 User s Guide 152 THIS PAGE INTENTIONALLY BLANK...
Page 154: ...BioFlo 4500 M1254 0050 User s Guide 154...
Page 160: ...BioFlo 4500 M1254 0050 User s Guide 160 THIS PAGE INTENTIONALLY BLANK...
Page 162: ...BioFlo 4500 M1254 0050 User s Guide 162 THIS PAGE INTENTIONALLY BLANK...
Page 164: ...BioFlo 4500 M1254 0050 User s Guide 164 THIS PAGE INTENTIONALLY BLANK...
Page 168: ...BioFlo 4500 M1254 0050 User s Guide 168 THIS PAGE INTENTIONALLY BLANK...
Page 186: ...BioFlo 4500 M1254 0050 User s Guide 186 THIS PAGE INTENTIONALLY BLANK...
Page 194: ...BioFlo 4500 M1254 0050 User s Guide 194 THIS PAGE INTENTIONALLY BLANK...
Page 196: ...BioFlo 4500 M1254 0050 User s Guide 196 THIS PAGE INTENTIONALLY BLANK...
Page 208: ...BioFlo 4500 M1254 0050 User s Guide 208 THIS PAGE INTENTIONALLY BLANK...
Page 214: ...BioFlo 4500 M1254 0050 User s Guide 214 Figure 49 Power Schematics Detail A...