BioFlo 4500 M1254-0050
User’s Guide
180
Figure 37: Addition Valve Detail
29.2
Addition & Inoculation
29.2.1 Method
A
1. Prepare a pre-sterilized additive in an appropriate addition vessel fitted
with tubing and a fitting that allows easy attachment to the quick-
connect.
2. Reduce the vessel pressure to 0 PSIG, manually or at the touchscreen.
3. Assuming a previously sterilized port, carefully remove the
inoculation port steam cap and aseptically attach an inoculation
container using silicone or Tygon type tubing.
4. To open the inoculation valve, press the black polymer handles in
(toward the vessel) and turn clockwise. This brings the valve’s
entrance port forward into the vessel.
5. Add the liquid inoculum by gravity feed or via a peristaltic pump.
6. After inoculation, close the inoculation valve: press the black polymer
handles in and turn counterclockwise, then release.
7. Break the addition vessel connection by removing the connecting
tubing from the quick-connect fitting.
ADDITION
VALVE
STEAM INLET
VALVE
TO
CONTAMINATED
CONDENSATE
DRAIN
CAP
PURE STEAM
Summary of Contents for BioFlo 4500
Page 2: ...BioFlo 4500 M1254 0050 User s Guide ii...
Page 4: ...BioFlo 4500 M1254 0050 User s Guide iv THIS PAGE INTENTIONALLY BLANK...
Page 6: ...BioFlo 4500 M1254 0050 User s Guide vi THIS PAGE INTENTIONALLY BLANK...
Page 8: ...BioFlo 4500 M1254 0050 User s Guide viii THIS PAGE INTENTIONALLY BLANK...
Page 10: ...BioFlo 4500 M1254 0050 User s Guide x THIS PAGE INTENTIONALLY BLANK...
Page 12: ...BioFlo 4500 M1254 0050 User s Guide xii THIS PAGE INTENTIONALLY BLANK...
Page 18: ...BioFlo 4500 M1254 0050 User s Guide xviii...
Page 34: ...BioFlo 4500 M1254 0050 User s Guide 34 THIS PAGE INTENTIONALLY BLANK...
Page 42: ...BioFlo 4500 M1254 0050 User s Guide 42 THIS PAGE INTENTIONALLY BLANK...
Page 60: ...BioFlo 4500 M1254 0050 User s Guide 60 THIS PAGE INTENTIONALLY BLANK...
Page 82: ...BioFlo 4500 M1254 0050 User s Guide 82 THIS PAGE INTENTIONALLY BLANK...
Page 88: ...BioFlo 4500 M1254 0050 User s Guide 88 THIS PAGE INTENTIONALLY BLANK...
Page 114: ...BioFlo 4500 M1254 0050 User s Guide 114 THIS PAGE INTENTIONALLY BLANK...
Page 138: ...BioFlo 4500 M1254 0050 User s Guide 138 THIS PAGE INTENTIONALLY BLANK...
Page 144: ...BioFlo 4500 M1254 0050 User s Guide 144 THIS PAGE INTENTIONALLY BLANK...
Page 152: ...BioFlo 4500 M1254 0050 User s Guide 152 THIS PAGE INTENTIONALLY BLANK...
Page 154: ...BioFlo 4500 M1254 0050 User s Guide 154...
Page 160: ...BioFlo 4500 M1254 0050 User s Guide 160 THIS PAGE INTENTIONALLY BLANK...
Page 162: ...BioFlo 4500 M1254 0050 User s Guide 162 THIS PAGE INTENTIONALLY BLANK...
Page 164: ...BioFlo 4500 M1254 0050 User s Guide 164 THIS PAGE INTENTIONALLY BLANK...
Page 168: ...BioFlo 4500 M1254 0050 User s Guide 168 THIS PAGE INTENTIONALLY BLANK...
Page 186: ...BioFlo 4500 M1254 0050 User s Guide 186 THIS PAGE INTENTIONALLY BLANK...
Page 194: ...BioFlo 4500 M1254 0050 User s Guide 194 THIS PAGE INTENTIONALLY BLANK...
Page 196: ...BioFlo 4500 M1254 0050 User s Guide 196 THIS PAGE INTENTIONALLY BLANK...
Page 208: ...BioFlo 4500 M1254 0050 User s Guide 208 THIS PAGE INTENTIONALLY BLANK...
Page 214: ...BioFlo 4500 M1254 0050 User s Guide 214 Figure 49 Power Schematics Detail A...