102
NATIONAL
INSTRUMENTS
MEASure:VOLTage?
Reads the measured output voltage. Returns a 5 digit string.
SCPI Command Syntax
MEASure:VOLTage?
GEN Command Syntax
MV?
Parameters
None
Default Suffix
V
Returned Parameters
<NR3>
MEASure:POWer?
Reads the measured output power. Returns a 5 digit string.
SCPI Command Syntax
MEASure:POWer?
GEN Command Syntax
MP?
Parameters
None
Default Suffix
W
Returned Parameters
<NR3>
7.12.6 DISPlay Subsystem
DISPlay:STATe
Turns front panel voltage and Current display toggle On or Off.
SCPI Command Syntax
DISPlay[:WINDow]:STATe <bool>
GEN Command Syntax
none
Parameters
0|OFF 1|ON
Examples
DISP:STAT 1 DISP:STAT OFF
Query Syntax
DISPlay[:WINDow]:STAT?
Returned Parameters
0 |1
DISPlay:FLASh
Makes front panel voltage and Current displays flash.
SCPI Command Syntax
DISPlay[:WINDow]:FLASh <bool>
GEN Command Syntax
none
Parameters
0|OFF 1|ON
Examples
DISP:FLASh 1 DISP:FLASh OFF