Wire size
AWG
Resistivity
Ohm/1000ft
Maximum length in feet to limit
voltage drop to 1V or less
5A
10A
20A
50A
80A
14
2.526
80
40
20
8
5
12
1.589
120
60
30
12
7.5
10
0.9994
200
100
50
20
12.5
8
0.6285
320
160
80
32
20
6
0.3953
500
250
125
50
31
4
0.2486
800
400
200
80
50
2
0.1564
1200
600
300
125
78
Table 3-2: Maximum wire length for 1V drop on lead (in feet)
Cross sect.
area
(mm
2
)
Resistivity
Ohm/Km
Maximum length in meters to limit
voltage drop to 1V or less
5A
10A
20A
50A
80A
2.5
8.21
24.0
12
6
2.4
1.5
4
5.09
39.2
18.6
9.8
4
2.5
6
3.39
59.0
29.4
14.8
5.8
3.6
10
1.95
102.6
51.2
25.6
10.2
6.3
16
1.24
160.0
80
40
16
10
25
0.795
250.0
125
62
25.2
15.7
35
0.565
354.0
177
88
35.4
22
Table 3-3: Maximum wire length for 1V drop on lead (in meters)
For currents not shown in Table 3-2 and 3-3, use the formula:
Maximum length=1000/(current x resistivity)
Where current is expressed in amperes and resistivity in ohms/km or ohms/1000ft.