115
NATIONAL
INSTRUMENTS
CHAPTER 8: ADVANCED FUNCTIONS
8.1 Introduction
This chapter describes the advanced functions of output programmable modes. The are three
programmable modes: FIX, WAVE and LIST. The user can program the output transient state. The
programmable modes are synchronized by input trigger (refer to section 8.5.1). According to
the selected mode, the power supply activates the output trigger on J3-3 (refer to section 8.5.2).
8.2 FIX Mode
Output value is changed via input trigger. Refer to 7.12.3 VOLTage:TRIGger and 7.12.4 CURRent:TRIGger.
Fig.8-1: Simplified Models of FIX Mode
Fig.8-2: Fix Mode Sequence example
0.01sec
Trig Initated
INIT:IMM
Trigger Event
Trigger Delay
TRIG:DEL 0.02
10V
5V
Set Voltage
VOLT:TRIG
10V
OUTPUT
5V
VOLTAGE
ABORT
INIT:CONT 1
0.02s