1-36
General information
ELECTRIC FIELD PROBE EP-408
Frequency range
1 MHz - 40 GHz
Level range
0.8 - 800 V/m
Overload
> 1000 V/m
Dynamic range
> 60 dB
Resolution
0.01 V/m
Sensitivity
0.8 V/m
Absolute error @ 200 MHz 6 V/m
±
0.8 dB
Flatness (1 MHz - 1 GHz)
±
1.5 dB
Flatness (1 - 3 GHz)
±
2 dB
Flatness (3 - 18 GHz)
±
2.5 dB
Flatness (18 - 26.5 GHz)
±
3 dB
Flatness (26.5 - 40 GHz)
±
4 dB
Isotropicity @ 200 MHz
±
0.8 dB (typical
±
0.5 dB @ 930 and 1800 MHz)
H-field rejection
> 20 dB
Temperature error
0.02 dB/°C
Calibration
internal into E
2
PROM
Size
317 mm length, 52 mm diameter
Weight
90 g
Fig. 1-29
EP-408 probe
Typical frequency response for EP-408 probe
-15