General information
1-47
ELECTRIC FIELD PROBE EP-201
Frequency range
60 MHz – 12 GHz
Level range
3 – 500 V/m
Overload
> 1000 V/m
Dynamic range
> 45 dB
Resolution
0.1 V/m
Sensitivity
8 V/m (instantaneous measurement with filter 10 Hz)
3 V/m (RMS or AVG 30 sec with filter 10 Hz)
Flatness @ 40 V/m
± 1.5 dB (150 MHz – 9.25 GHz)
± 3 dB (60 MHz – 12 GHz)
Isotropicity @ 40 V/m @ 200 MHz
± 0.6 dB
H-field rejection
> 20 dB
A/D Conversion
one converter for every axis
Calibration
On board EEPROM
Microcontroller
On board
Volume sensor
3 mm diameter sphere
Size tube
180mm length x 4 mm diameter
Size
300 mm length x 18 mm diameter
Weight
85 g
Typical frequency response for EP201 probe
-4
-3
-2
-1
0
1
2
3
4
10
100
1000
10000
100000
MHz
dB