1-44
General information
ELECTRIC FIELD PROBE EP-33C
Frequency range
2110 MHz – 2170 MHz
Level range
0.03 – 30 V/m
Overload
> 120 V/m
Dynamic range
> 60 dB
Resolution
0.001 V/m
Sensitivity
0.03 V/m
Absolute error @ 2140 MHz and 2 V/m
±
1 dB
Flatness (2110 - 2170 MHz)
+ 0.2 dB / -1.8 dB
OFF Band attenuation respect to 2140 MHz
1880 MHz
> 10 dB
2320 MHz
> 10 dB
Isotropicity
±
0.8 dB (typical
±
0.5 dB)
Rejection to H field
> 20 dB
Temperature error
0°C ÷ 60°C =
±
0.2dB
-20°C ÷ 0°C = -0.1 dB/°C
Drift Frequency Vs Temperature
40°C ÷ 60°C =
±
100 kHz
-20°C ÷ 40°C = -100 kHz/°C
Calibration
E
2
PROM internal
Size
317 mm length, 58 mm diameter
Weight
100 g
T ypi c al fr eq uen cy r es pons e for E P 33C probe
- 20
- 18
- 16
- 14
- 12
- 10
-8
-6
-4
-2
0
1679.4
1863.6
2047.9
2232.1
2416.4
[M H z]
[d B]