NanoPhotometer
®
NP80/N60/N50/C40
User Guide Version 3.1
73
cf
Correction factor for spectrophotometer
multiplier
Multiplier of sample
M
ORE
A
PPS
The More Apps icon located on the home screen opens another menu screen with access to
icons for additional applications available on the NanoPhotometer
®
. The applications featured in
this menu include: wavelength, concentration, wavescan, absorbance/ratio, standard curve and
custom applications.
M
ORE
A
PPS
:
W
AVELENGTH
M
ETHOD
O
VERVIEW
In the wavelength application it is possible to measure simple absorbance (A) and %
transmittance (%Trans. / only in cuvette mode) of a sample at specific wavelengths. It is
possible to add up to 20 different wavelengths.
M
EASUREMENT
P
ROTOCOL
1. Select the More Apps icon from the home screen and the
Wavelength icon from the More Apps screen.
2.
To change between NanoVolume and Cuvette application, use the
Change to Cuvette/NanoVolume button below the parameter area
(NP80/C40 only).
3.
For NanoVolume application:
Select the dilution depending on the sample concentration.
Note
: There is no automatic path length setting in this method. Select either a virtual
dilution of 15 (path length 0.67 mm) or of 140 (path length 0.07 mm).
For cuvette application:
Select the path length depending on the used cuvette.
Options are: 0.5 mm, 1 mm, 2 mm, 5 mm and 10 mm