NanoPhotometer
®
NP80/N60/N50/C40
User Guide Version 3.1
108
No connection possible
Connection is lost between control devices and
NanoPhotometer
®
. Check the connection. If the
message still appears please create a log file and
send it to Implen Customer Support.
Server Not Available
Internal software communication problem during
an open measurement session. Reboot the
NanoPhotometer
®
. If the message still appears
please create a log file and send it to Implen
Customer Support.
Server Side Failure
Internal software communication problem during
an open measurement session. Reboot the
NanoPhotometer
®
. If the message still appears
please create a log file and send it to Implen
Customer Support.
Software update failed. Please contact Implen
customer support.
Make sure that the sample name is NPOS.bin
without any additions or extensions like from
multiple downloads, otherwise rename file to be
NPOS.bin exactly. NPOS.bin file must be in the
root folder of the USB flash drive. Try the update a
second time, if it still fails please create a log file
and send it to Implen Customer Support.
Stored method/result was created for unavailable
mode NanoVolume/Cuvette/Submicroliter Cell!
Message appears if stored results or a stored
method made on another NanoPhotometer
®
type is
opened with an unavailable mode like Cuvette or
NanoVolume e.g. data or methods made in cuvette
mode and opened on a N60 (NanoVolume only).
Cuvette port blocked - please check.
The automatic sliding cover of the
NanoPhotometer
®
NP80 could not close properly.
Please check if the cuvette port is empty. If
message still appears please contact Implen
Customer Support.