message: “Do you want to delete all/selected files?” selecting cancel (x) will return to the delete
menu screen and confirming with delete, deletes the selected data.
Note
: The delete function is not available for the software version designed for smartphones.
B
ASIC
O
PERATION
The NanoPhotometer
®
product line offers a complete solution for nanovolume
(N50/N60/NP80/C40) and standard volume (C40/NP80) applications. With the
NanoPhotometer
®
nanovolume applications the sample volume ranges from 0.3 µl to a
maximum sample volume of 2.0 µl. Standard Cuvette applications can be performed with 10
mm, 5 mm, 2 mm, 1 mm and 0.5 mm path length quartz, glass, or plastic cuvettes with a center
height of 8.5 cm.
Note
: NanoVolume application in the NanoPhotometer
®
C40 is only available with the optional
accessory “submicroliter cell” from Implen.
N
ANO
V
OLUME
M
EASUREMENT
B
ASICS
(N50/N60/NP80)
1. Select a method depending on your sample and set the parameters for the measurement.
2. Ensure that the sample window and the mirror in the sample arm are clean.
3. Use the integrated vortexer (N60 and NP80 only) to mix your sample to achieve a
homogenous sample
4. Raise the sample arm and pipette the appropriate amount of blank solution onto the
illuminated sample window. The Illumination automatically turns off when the arm is
lowered.