204
Sensor Data
GB
SUPREMATouch
Start-up
Preadjustment required before first calibration and when changing sensor
Preadjustment
Connect the digital multimeter to the test sockets on the MAI module.
Bridge current
setting:
310 mA
Zero adjustment
by zero gas:
Zero setting Ua = 400–450 mV
Sensitivity adjust-
ment with
measuring gas
Measuring range level Ua = 1950–2100 mV
or with the value of the existing gas concentration
according to:
Ua (mV) = C / 100 * 1600 + 400
C = Span gas concentration in % of measuring range
Warm-up period
15 min minutes for preadjustment,
2 hours for calibration
Function test
Span gas application via:
• Test cap with 1.0 l/ min
Calibration
Calibration procedure according to chapter 7 "Calibration".
For approved gas types, measuring ranges, lower alarm levels and condi-
tions for calibration see list of components (Order No.: D0792420)
Possible other gas types and measuring ranges on request.
Open or Short Circuit Fault Indication:
X=
Signal fail (FAIL-LED)
XX=
Alarm LEDS, Signal exceeded, Signal fail (FAIL-LED)
XXX=
only alarms
XXXX=
no influence on operation
Open-circuit
at the MAT
(TS) Module
Wire
-X1/1
Wire
-X1/2
Wire
-X1/3
Wire
-X1/4
Wire
-X1/5
Disconnect plug of
MAT (TS)
Failure
indication
XX
X
X
X
XX
X
Open-circuit
at max. cable
length
Wire
-X1/1
Wire
-X1/2
Wire
-X1/3
Wire
-X1/4
Wire
-X1/5
Disconnect plug of
MAT (TS)
Failure
indication
XX
X
X
X
XX
X
Short-circuit
at the MAT
(TS) Module
Wire
-X1/1/-
X1/2
Wire
-X1/1/-
X1/3
Wire
-X1/1/-
X1/4
Wire
-X1/1/-
X1/5
Wire
-X1/2/-
X1/3
Wire
-X1/2/-
X1/4
Wire
-X1/2/-
X1/5
Wire
-X1/3/-
X1/4
Wire
-X1/3/-
X1/5
Wire
-X1/4/-
X1/5
Failure
indication
X
XX
XX
X
XX
X
X
X
X
XX
Short-circuit
at max. cable
length
Wire
-X1/1/-
X1/2
Wire
-X1/1/-
X1/3
Wire
-X1/1/-
X1/4
Wire
-X1/1/-
X1/5
Wire
-X1/2/-
X1/3
Wire
-X1/2/-
X1/4
Wire
-X1/2/-
X1/5
Wire
-X1/3/-
X1/4
Wire
-X1/3/-
X1/5
Wire
-X1/4/-
X1/5
Failure
indication
XXXX XX
X
X
XX
X
X
X
X
XXXX
Summary of Contents for SUPREMATouch MCP 20
Page 1: ...MSAsafety com Operating Manual SUPREMATouch Fire and Gas Warning Unit Order No 10121863 05...
Page 8: ...MSAsafety com User Instruction Manual SUPREMATouch Fire and Gas Warning Unit...
Page 72: ...MSAsafety com Service and Maintenance Guide SUPREMATouch Fire and Gas Warning Unit...
Page 232: ...232 Dimensions GB SUPREMATouch 18 Dimensions 18 1 Rack...
Page 233: ...233 Dimensions GB SUPREMATouch 18 2 Rail mounted Modules MRO8 TS Module MRO16 TS Module...
Page 235: ...235 Dimensions GB SUPREMATouch MRO20 8 TS Module 1 3 2 69 90...
Page 236: ...236 Dimensions GB SUPREMATouch MRO20 16 TS Module 2 5 2 64 73 relay dependent 90...
Page 237: ...237 Dimensions GB SUPREMATouch MRC TS Module MGT40 TS Module...
Page 238: ...238 Dimensions GB SUPREMATouch MHD TS Module MAT TS Module...
Page 241: ...For local MSA contacts please visit us at MSAsafety com Because every life has a purpose...